Advance-Esthetic LLC
Fort Lauderdale, FL, US
ADVANCE-ESTHETIC LLC appears in FDA device records in Fort Lauderdale, FL. FDA records list no 510(k) clearances, no PMA records and one registered establishment. As of , FDA records show one recall for this company, and at least one establishment registration is current.
FDA records held for this company
510(k) clearance history
No 510(k) clearance records for this company in the ingested FDA data.
No 510(k) clearance records in the ingested FDA data.
Registered establishments
| Registration | FEI | Establishment | Address | Registration expires | Product codes listed |
|---|---|---|---|---|---|
| 3013397886 | 3013397886 | ADVANCE-ESTHETIC LLC | 4101 Ravenswood Rd Ste 110, Fort Lauderdale, FL 33312 US | 2026 |
Registration numbers link to the FDA registration and listing record for that establishment.
FDA product codes
- GFEBrush, Dermabrasion, Powered
- KNMDevice, Pressure Applying
- HCJDevice, Skin Potential Measurement
- ILYLamp, Infrared, Therapeutic Heating
- OHSLight Based Over The Counter Wrinkle Reduction
- GMBLight, Wood'S, Fluorescence
- ISAMassager, Therapeutic, Electric
- OLPOver-The-Counter Powered Light Based Laser For Acne
- GEXPowered Laser Surgical Instrument
- ONFPowered Light Based Non-Laser Surgical Instrument With Thermal Effect
Recall records
| Recall | Product | Reason | Classification | Initiated | Status |
|---|---|---|---|---|---|
| Z-0095-2024 | MED-810A Zemits NDPrime Laser products | Advance-Esthetic LLC failed to comply with the applicable standards regarding Electronic Product Radiation Control (EPRC) regulations associated with the laser product performance standards and certification. This Recall provides correction by the firm regarding Zemits NdPrime Yag Tattoo Removal Laser. The product requires update or is missing the following in part: remote interlock connector, safety interlocks, manual reset mechanism, beam attenuator, tests for determination of compliance, certification, class IV designation and warning, labels, and labeling. | Class II | Open, Classified |
Verification record
There is no verification record for this company. Most companies on this site have none: a record exists only where a company asked to be verified and MDM ran the checks, so the absence of one says nothing about the company.
Where there is a record, this is where it appears: what was checked, when it was checked, and the current status. An MDM Verified badge on another website links to a verify address on this domain that resolves to this section. A badge that does not resolve here is not ours.
MDM coverage
MDM coverage of this company lists here, newest first. Submit news.
Is this your company?
This page is built from public FDA records. Company pages are free and factual, and stay free whether or not anyone claims them. Membership never changes editorial coverage or vetting outcomes.
Claiming lets Advance-Esthetic LLC edit what it says about itself and list the categories it works in. It grants no badge: claiming is not verification.
Data as of , FDA export datePage updated Source: openFDA, public domain