Beyond Laser Systems, LLC
Austin, TX, US
Beyond Laser Systems, LLC appears in FDA device records in Austin, TX. FDA records list no 510(k) clearances, no PMA records and no registered establishments. As of , FDA records show 3 recalls for this company.
FDA records held for this company
- 510(k) clearances
- 0
- PMA records
- 0
- Recall records
- 3
- Registered establishments
- 0
510(k) clearance history
No 510(k) clearance records for this company in the ingested FDA data.
No 510(k) clearance records in the ingested FDA data.
Registered establishments
No FDA establishment registrations recorded against this company.
FDA product codes
No FDA product codes recorded against this company.
Recall records
| Recall | Product | Reason | Classification | Initiated | Status |
|---|---|---|---|---|---|
| Z-2113-2025 | Laser light show projectors manufactured by BLS, Model PURE 5000/10000. | The remote interlock connector, which is required by 21 CFR 1040.10(f)(3), in the remote stop box was not installed the connector or was mechanically installed but not wired. The labels placed on the projectors do not meet requirements specified in 21 CFR 1010.4 (Variances), 21 CFR 1010.2 (Certification), 21 CFR 1010.3 (Identification) and 21 CFR 1040.10 (g)(Labelling requirements). The identification label does not include the date of manufacture as required by 21 CFR 1010.3. The corrective action plan covers three families of laser light show of projectors: DIODE. CLUB and PURE with three different hardware configurations such as the number of the number of RJ45 ports, radiant power of emissions and network configurations. No specific user manuals were provided to each model family as required by 21 CFR 1040.10(h)(1) (Informational requirements (1) User information). The Quality Control test did not validate the manual reset function to enable resumption of laser radiation emission from Class IV laser projectors after an interruption caused using a remote interlock. No supplemental product reports, as required according to 21 CFR 1002.11. | Class II | Open, Classified | |
| Z-2112-2025 | Laser light show projectors manufactured by BLS, Model CLUB 1000/2000/4000. | The remote interlock connector, which is required by 21 CFR 1040.10(f)(3), in the remote stop box was not installed the connector or was mechanically installed but not wired. The labels placed on the projectors do not meet requirements specified in 21 CFR 1010.4 (Variances), 21 CFR 1010.2 (Certification), 21 CFR 1010.3 (Identification) and 21 CFR 1040.10 (g)(Labelling requirements). The identification label does not include the date of manufacture as required by 21 CFR 1010.3. The corrective action plan covers three families of laser light show of projectors: DIODE. CLUB and PURE with three different hardware configurations such as the number of the number of RJ45 ports, radiant power of emissions and network configurations. No specific user manuals were provided to each model family as required by 21 CFR 1040.10(h)(1) (Informational requirements (1) User information). The Quality Control test did not validate the manual reset function to enable resumption of laser radiation emission from Class IV laser projectors after an interruption caused using a remote interlock. No supplemental product reports, as required according to 21 CFR 1002.11. | Class II | Open, Classified | |
| Z-2111-2025 | Laser light show projectors manufactured by BLS, Model DIODE 1000/2000/3000. | The remote interlock connector, which is required by 21 CFR 1040.10(f)(3), in the remote stop box was not installed the connector or was mechanically installed but not wired. The labels placed on the projectors do not meet requirements specified in 21 CFR 1010.4 (Variances), 21 CFR 1010.2 (Certification), 21 CFR 1010.3 (Identification) and 21 CFR 1040.10 (g)(Labelling requirements). The identification label does not include the date of manufacture as required by 21 CFR 1010.3. The corrective action plan covers three families of laser light show of projectors: DIODE. CLUB and PURE with three different hardware configurations such as the number of the number of RJ45 ports, radiant power of emissions and network configurations. No specific user manuals were provided to each model family as required by 21 CFR 1040.10(h)(1) (Informational requirements (1) User information). The Quality Control test did not validate the manual reset function to enable resumption of laser radiation emission from Class IV laser projectors after an interruption caused using a remote interlock. No supplemental product reports, as required according to 21 CFR 1002.11. | Class II | Open, Classified |
Verification record
There is no verification record for this company. Most companies on this site have none: a record exists only where a company asked to be verified and MDM ran the checks, so the absence of one says nothing about the company.
Where there is a record, this is where it appears: what was checked, when it was checked, and the current status. An MDM Verified badge on another website links to a verify address on this domain that resolves to this section. A badge that does not resolve here is not ours.
MDM coverage
MDM coverage of this company lists here, newest first. Submit news.
Is this your company?
This page is built from public FDA records. Company pages are free and factual, and stay free whether or not anyone claims them. Membership never changes editorial coverage or vetting outcomes.
Claiming lets Beyond Laser Systems, LLC edit what it says about itself and list the categories it works in. It grants no badge: claiming is not verification.
Data as of , FDA export datePage updated Source: openFDA, public domain