CME America, LLC

Golden, CO, US

CME America, LLC appears in FDA device records in Golden, CO. FDA records list no 510(k) clearances, no PMA records and no registered establishments. As of , FDA records show 48 recalls for this company.

FDA records held for this company

510(k) clearances
0
PMA records
0
Recall records
48
Registered establishments
0

510(k) clearance history

No 510(k) clearance records for this company in the ingested FDA data.

No 510(k) clearance records in the ingested FDA data.

Registered establishments

No FDA establishment registrations recorded against this company.

FDA product codes

No FDA product codes recorded against this company.

Recall records

Showing 25 of 48
Device recall records naming this company, most recent first
RecallProductReasonClassificationInitiatedStatus
Z-0530-2021CME America T-Syringe Pump - T34 - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.Class IITerminated
Z-0529-2021CME America T-Syringe Pump - Ameritus 1st Enteral Pump - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.Class IITerminated
Z-0528-2021CME America T-Syringe Pump - Neofeed - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.Class IITerminated
Z-0527-2021CME America T-Syringe Pump - T34L PCA - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.Class IITerminated
Z-0526-2021CME America T-Syringe Pump - NeoThrive - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.Class IITerminated
Z-0525-2021CME America T-Syringe Pump -T34L - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.Class IITerminated
Z-2589-2020BodyGuard Microset w/ 0.2 mic filter and lower y-site with female luer, REF: A120-160CYFRVInfusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapyClass ITerminated
Z-2588-2020BodyGuard Microset With Female Luer, REF: A120-160XPSNInfusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapyClass ITerminated
Z-2587-2020CMExpress Needleless Y Site Microbore Set, REF: A120-160CYNVKInfusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapyClass ITerminated
Z-2586-2020CMExpress Microbore Set, REF: A120-160CInfusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapyClass ITerminated
Z-2585-2020Standard BodySet with Needleless Connectors, REF: A120-125XSInfusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapyClass ITerminated
Z-2584-2020BodyGuard Microset w/ 0.2 mic filter and lower y-site with female luer, REF: A120-160XYBFInfusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapyClass ITerminated
Z-2583-2020BodyGuard Microset with Filter and Manual Priming Valve, REF: A120-112XSFVInfusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapyClass ITerminated
Z-2582-2020BodyGuard set with Female Luer, REF: A120-003XSNInfusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapyClass ITerminated
Z-2581-2020BodyGuard Microset with Male Luer Connectors, REF: A100-163XEBFInfusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapyClass ITerminated
Z-2580-2020CMExpress Microbore Sets, REF: A120-161CInfusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapyClass ITerminated
Z-2579-2020CMExpress Microbore Sets, REF: A120-160C2YRVInfusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapyClass ITerminated
Z-2578-2020CMExpress Microbore Sets, REF: A120-160CYFRVInfusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapyClass ITerminated
Z-2577-2020BodyGuard Microset, REF: A120-160GCFAInfusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapyClass ITerminated
Z-2576-2020BodyGuard Microset, REF: A120-125XSEInfusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapyClass ITerminated
Z-2575-2020BodyGuard Microset, REF: A120-003XS2YVInfusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapyClass ITerminated
Z-2574-2020BodyGuard Microset, REF: A100-163XSLInfusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapyClass ITerminated
Z-2573-2020BodyGuard Microset, REF: A100-163XE90SInfusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapyClass ITerminated
Z-2572-2020BodyGuard Microset with Filter, REF: A100-164XESFInfusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapyClass ITerminated
Z-2571-2020BodyGuard Microset with Filter, REF: A120-160XSFEInfusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapyClass ITerminated

Verification record

There is no verification record for this company. Most companies on this site have none: a record exists only where a company asked to be verified and MDM ran the checks, so the absence of one says nothing about the company.

Where there is a record, this is where it appears: what was checked, when it was checked, and the current status. An MDM Verified badge on another website links to a verify address on this domain that resolves to this section. A badge that does not resolve here is not ours.

What MDM checks, and what it does not

MDM coverage

MDM coverage of this company lists here, newest first. Submit news.

Is this your company?

This page is built from public FDA records. Company pages are free and factual, and stay free whether or not anyone claims them. Membership never changes editorial coverage or vetting outcomes.

Claiming lets CME America, LLC edit what it says about itself and list the categories it works in. It grants no badge: claiming is not verification.

Claim this page

Data as of , FDA export datePage updated Source: openFDA, public domain