Conmed Linvatec Endoscopy Division
Santa Barbara, CA
Conmed Linvatec Endoscopy Division appears in FDA device records in Santa Barbara, CA. FDA records list no 510(k) clearances, no PMA records and no registered establishments. As of , FDA records show 2 recalls for this company.
FDA records held for this company
- 510(k) clearances
- 0
- PMA records
- 0
- Recall records
- 2
- Registered establishments
- 0
FDA records under these names
1 nameFDA records resolved to Conmed Linvatec Endoscopy Division carry these names as well, as FDA wrote them. Each is the firm named on a registration, a 510(k), a PMA, a recall or an enforcement report that was joined to this company on a shared FDA identifier or on a matching name.
- Linvatec Imagingrecall
510(k) clearance history
No 510(k) clearance records for this company in the ingested FDA data.
No 510(k) clearance records in the ingested FDA data.
Registered establishments
No FDA establishment registrations recorded against this company.
FDA product codes
No FDA product codes recorded against this company.
Recall records
| Recall | Product | Reason | Classification | Initiated | Status |
|---|---|---|---|---|---|
| Z-2458-2008 | ConMed Linvatec Surgical Video Cart, Catalogue Number: VP8500 | The recall was initiated because ConMed Linvatec has determined there is a possibility the wheel caster(s) may become loose which may result in the wheel caster(s) falling out of the base of the VP8500 Video Cart. The potential for the wheel caster or casters to fall out of the base of the cart may result in a potential tip hazard. | Terminated | ||
| Z-1531-05 | Linvatec LS 7500, 300 Watt Xenon Light Source Catalog Numbers: LS7500, LS7500L (loaner), LS7500R (refurbished) | Through the complaint handling system, the company has determine there are performance issued involving the LS7500, 300 Watt Xenon Light Source. The performance issues result in a low light output that can create poor illumination of the surgical site. | Terminated |
Verification record
There is no verification record for this company. Most companies on this site have none: a record exists only where a company asked to be verified and MDM ran the checks, so the absence of one says nothing about the company.
Where there is a record, this is where it appears: what was checked, when it was checked, and the current status. An MDM Verified badge on another website links to a verify address on this domain that resolves to this section. A badge that does not resolve here is not ours.
MDM coverage
MDM coverage of this company lists here, newest first. Submit news.