Epix Therapeutics, Inc
Sunnyvale, CA, US
EPIX THERAPEUTICS, INC appears in FDA device records in Sunnyvale, CA. FDA records list no 510(k) clearances, no PMA records and no registered establishments. As of , FDA records show 6 recalls for this company.
FDA records held for this company
- 510(k) clearances
- 0
- PMA records
- 0
- Recall records
- 6
- Registered establishments
- 0
510(k) clearance history
No 510(k) clearance records for this company in the ingested FDA data.
No 510(k) clearance records in the ingested FDA data.
Registered establishments
No FDA establishment registrations recorded against this company.
FDA product codes
No FDA product codes recorded against this company.
Recall records
| Recall | Product | Reason | Classification | Initiated | Status |
|---|---|---|---|---|---|
| Z-1322-2021 | EPIX Therapeutics DiamondTemp Unidirectional Ablation Catheter, Large Curve, 7.5 French Size(2.48 mm), Connector Type: ODU 19 PIN, REF:DT200L, SterileEO, Rx Only, UDI: 00812499030013 | It has been identified that the catheter size on the packaging labeling does not correspond to the widest part of the catheter. In addition the IFU contains the incorrect diameter of the catheter. This could result in blood stagnation, clots or air emboli. | Class II | Open, Classified | |
| Z-1321-2021 | EPIX Therapeutics DiamondTemp Unidirectional Ablation Catheter, Small Curve, 7.5 French Size(2.48 mm), Connector Type: ODU 19 PIN, REF:DT100S,SterileEO, Rx Only, UDI: 00812499030006 | It has been identified that the catheter size on the packaging labeling does not correspond to the widest part of the catheter. In addition the IFU contains the incorrect diameter of the catheter. This could result in blood stagnation, clots or air emboli. | Class II | Open, Classified | |
| Z-1320-2021 | EPIX Therapeutics DiamondTemp Bidrectional Large Curve Ablation Catheter, 7.5 French Size(2.48 mm), Connector Type: ODU 19 PIN, REF:CEDTB400L, SterileEO, UDI: 00812499030303 | It has been identified that the catheter size on the packaging labeling does not correspond to the widest part of the catheter. In addition the IFU contains the incorrect diameter of the catheter. This could result in blood stagnation, clots or air emboli. | Class II | Open, Classified | |
| Z-1319-2021 | EPIX Therapeutics DiamondTemp Bidrectional Small Curve Ablation Catheter, 7.5 French Size(2.48 mm), Connector Type: ODU 19 PIN, REF:CEDTB300S, SterileEO, UDI: 00812499030358 | It has been identified that the catheter size on the packaging labeling does not correspond to the widest part of the catheter. In addition the IFU contains the incorrect diameter of the catheter. This could result in blood stagnation, clots or air emboli. | Class II | Open, Classified | |
| Z-1318-2021 | EPIX Therapeutics DiamondTemp Unidirectional Large Curve Ablation Catheter, 7.5 French Size(2.48 mm), Connector Type: ODU 19 PIN, REF: CEDT200L, Sterile EO, UDI: 00812499030259 | It has been identified that the catheter size on the packaging labeling does not correspond to the widest part of the catheter. In addition the IFU contains the incorrect diameter of the catheter. This could result in blood stagnation, clots or air emboli. | Class II | Open, Classified | |
| Z-1317-2021 | EPIX Therapeutics DiamondTemp Unidirectional Small Curve Ablation Catheter, 7.5 French Size(2.48 mm), Connector Type: ODU 19 PIN, REF CEDT100S, SterileEO, UDI 00812499030198 | It has been identified that the catheter size on the packaging labeling does not correspond to the widest part of the catheter. In addition the IFU contains the incorrect diameter of the catheter. This could result in blood stagnation, clots or air emboli. | Class II | Open, Classified |
Verification record
There is no verification record for this company. Most companies on this site have none: a record exists only where a company asked to be verified and MDM ran the checks, so the absence of one says nothing about the company.
Where there is a record, this is where it appears: what was checked, when it was checked, and the current status. An MDM Verified badge on another website links to a verify address on this domain that resolves to this section. A badge that does not resolve here is not ours.
MDM coverage
MDM coverage of this company lists here, newest first. Submit news.
Is this your company?
This page is built from public FDA records. Company pages are free and factual, and stay free whether or not anyone claims them. Membership never changes editorial coverage or vetting outcomes.
Claiming lets Epix Therapeutics, Inc edit what it says about itself and list the categories it works in. It grants no badge: claiming is not verification.
Data as of , FDA export datePage updated Source: openFDA, public domain