H & P Industries, Inc. dba Triad Group
Hartland, WI
H & P Industries, Inc. dba Triad Group appears in FDA device records in Hartland, WI. FDA records list no 510(k) clearances, no PMA records and no registered establishments. As of , FDA records show one recall for this company.
FDA records held for this company
- 510(k) clearances
- 0
- PMA records
- 0
- Recall records
- 1
- Registered establishments
- 0
510(k) clearance history
No 510(k) clearance records for this company in the ingested FDA data.
No 510(k) clearance records in the ingested FDA data.
Registered establishments
No FDA establishment registrations recorded against this company.
FDA product codes
No FDA product codes recorded against this company.
Recall records
| Recall | Product | Reason | Classification | Initiated | Status |
|---|---|---|---|---|---|
| Z-0934-2011 | Lubricating Jelly, Sterile, Non-staining, Made in the USA, packaged into the following size containers 5 gram tube, 2 ounce and 4 ounce tubes, 4 ounce bottle, 3 and 5 gram packets Intended use: A patient lubricant is a device intended for medical purposes that is used to lubricate a body orifice to facilitate entry of a diagnostic or therapeutic device. Brand names: 1. Allegiance, Distributed by: Cardinal Health, McGaw Park, IL 60085-6787, USA, Cat. LJT2. Net Wt. 2 oz., Cat. LJF3, Net Wt. 3g, Cat. LJT4, Net Wt. 4 oz., Cat. LJT5, Net Wt. 5 g; 2. Select medical products, manufactured for PSS World Medical, Inc., 4345 Southpoint Blvd., Jacksonville, FL 32216, Reorder #137, 3 g/packet, Reorder #136, 4 oz.; 3. Novaplus, Manufactured for Novation, Inc., Irving, TX 75039, Manufactured by: Triad Group, Inc., 700 West North Shore Drive, Hartland, WI 53029, Cat. No. V10-8344, Net Wt. 3 grams, Cat. No. V10-8919, Net Wt. 4 oz., Cat. No. V10-8917, Net Wt. 2 oz.; 4. Triad Group, Inc., 700 West North Shore Drive, Hartland, WI 53029, Cat. No. 10-8917, Net Wt. 2 oz, Cat. No. 10-8946, Net Wt. 5 g, Triad Plus Cat. No. 11-8344, 3 g., Cat. No. 11-8472, 5 grams, Cat. No. 10-8919, Net Wt. 4 oz, Cat. No. 10-8500, Net Wt. 4 oz; 5. IMCO, Manufactured for: Independent Medical Co-op, Inc., Daytona Beach, FL 32114. Reorder No. 8919-IMC Net Wt. 4 oz.; 6. McKesson Medi-Pak Performance, Marketed by McKesson Corporation, McKesson Surgical, Richmond, VA 23228. Reorder No. 66-8919, Net Wt. 4 oz.; 7. Henry Schein, Distributed by: Henry Schein Inc., Melville, NY 11747. 104-9637, Net Wt. 4 fl. oz. | Triad Group is recalling all lots of Sterile Lubricating Jelly manufactured by Triad Group. This recall has been initiated due to concerns expressed by the Food and Drug Administration regarding the validation of the gamma radiation sterilization cycles for these products. | Terminated |
Verification record
There is no verification record for this company. Most companies on this site have none: a record exists only where a company asked to be verified and MDM ran the checks, so the absence of one says nothing about the company.
Where there is a record, this is where it appears: what was checked, when it was checked, and the current status. An MDM Verified badge on another website links to a verify address on this domain that resolves to this section. A badge that does not resolve here is not ours.
MDM coverage
MDM coverage of this company lists here, newest first. Submit news.
Is this your company?
This page is built from public FDA records. Company pages are free and factual, and stay free whether or not anyone claims them. Membership never changes editorial coverage or vetting outcomes.
Claiming lets H & P Industries, Inc. dba Triad Group edit what it says about itself and list the categories it works in. It grants no badge: claiming is not verification.
Data as of , FDA export datePage updated Source: openFDA, public domain