Medical Positioning Inc

Lenexa, KS, US

MEDICAL POSITIONING, INC. appears in FDA device records in Lenexa, KS. FDA records list 3 510(k) clearances between and , no PMA records and one registered establishment. As of , FDA records show 9 recalls for this company, and at least one establishment registration is current.

FDA records held for this company

510(k) clearances
3
PMA records
0
Recall records
9
Registered establishments
1

510(k) clearance history

FDA records hold 3 510(k) clearances for Medical Positioning Inc between 2001 and 2009. The busiest year on record is 2001, with 1. The most recent is 2009, with 1.

Clearances by year, as a table
510(k) clearance decisions for Medical Positioning Inc, by decision year
YearClearances
20011
20021
20091
510(k) clearances on record, most recent first
K numberDeviceProduct codeDecisionDecision date
K082967MORTARA STRESS ECHO BEDDPSSubstantially Equivalent
K021171Q-STRESS ECHO BEDDPSSubstantially Equivalent
K002822VERTEX SYSTEMDPSSubstantially Equivalent

Registered establishments

FDA registered establishments for Medical Positioning Inc, with the product codes listed at each
RegistrationFEIEstablishmentAddressRegistration expiresProduct codes listed
19320561000118382MEDICAL POSITIONING, INC.9738 Pflumm Road, Lenexa, KS 66215 US2026

Registration numbers link to the FDA registration and listing record for that establishment.

FDA product codes

Recall records

Device recall records naming this company, most recent first
RecallProductReasonClassificationInitiatedStatus
Z-2211-2014Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1219. Product Usage: Used to conduct syncope examinations.Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction.Class IITerminated
Z-2210-2014Medical Positioning Rapid Response HUT table with Dual 2 Way Drop Section, Powered Medical Examination table, Model # 1217. Product Usage: Used to conduct syncope examinations.Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction.Class IITerminated
Z-2209-2014Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1019. Product Usage: Used to conduct syncope examinations.Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction.Class IITerminated
Z-2208-2014Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1017. Product Usage: Used to conduct syncope examinations.Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction.Class IITerminated
Z-1211-2012MPI Rapid Response HUT Table, Model #RRHUT 1017. The firm name on the label is Medical Positioning Inc., Kansas City, MO. Designed for multiple uses including a conventional examination table, unexplained syncope, neurocardiogenic syncope, chronic fatigue syndrome, and most noninvasive procedures.The warning label regarding electrical shock was not placed on the device warning to have the power switch in the "OFF" position prior to connecting the unit to electric power.Terminated
Z-1210-2012MPI Rapid Response HUT Table, Model #RRHUT 1217 with Dual 2 Way Drop Section. The firm name on the label is Medical Positioning Inc., Kansas City, MO. Designed for multiple uses including a conventional examination table, unexplained syncope, neurocardiogenic syncope, chronic fatigue syndrome, and most noninvasive procedures. With the optional drop section, the use expands to echo/stress echo/Doppler studies, and vascular and carotid echo.The warning label regarding electrical shock was not placed on the device warning to have the power switch in the "OFF" position prior to connecting the unit to electric power.Terminated
Z-1975-2011Rapid Response HUT (Head Up Tilt) Table. Model 1217. Powered electrophysiology tilt table with dual 2 way drop section. Table that can tilt -15 degree to 90 degree (Trendelenburg). Medical Positioning Incorporated. 800-593-3246. Used in clinical and hospital settings by trained radiologists.The tilt tables work intermittently and some have become stuck in an inclined or reclined position during procedures.Terminated
Z-1974-2011Rapid Response HUT (Head Up Tilt) Table. Model 1117. Powered electrophysiology tilt table with single drop section. Table that can tilt -15 degrees to 90 degrees (Trendelenburg). Medical Positioning Incorporated.800-593-3246. Used in clinical and hospital settings by trained radiologists.The tilt tables work intermittently and some have become stuck in an inclined or reclined position during procedures.Terminated
Z-1973-2011Rapid Response H.U.T. (Head Up Tilt) Table. Model 1017. Powered electrophysiology tilt table. Table can tilt -15 degrees to 90 degrees (Trendelenburg). Medical Positioning Incorporated.800-593-3246. Used in clinical and hospital settings by trained radiologists.The tilt tables work intermittently and some have become stuck in an inclined or reclined position during procedures.Terminated

Verification record

There is no verification record for this company. Most companies on this site have none: a record exists only where a company asked to be verified and MDM ran the checks, so the absence of one says nothing about the company.

Where there is a record, this is where it appears: what was checked, when it was checked, and the current status. An MDM Verified badge on another website links to a verify address on this domain that resolves to this section. A badge that does not resolve here is not ours.

What MDM checks, and what it does not

MDM coverage

MDM coverage of this company lists here, newest first. Submit news.

Is this your company?

This page is built from public FDA records. Company pages are free and factual, and stay free whether or not anyone claims them. Membership never changes editorial coverage or vetting outcomes.

Claiming lets Medical Positioning Inc edit what it says about itself and list the categories it works in. It grants no badge: claiming is not verification.

Claim this page

Data as of , FDA export datePage updated Source: openFDA, public domain