Zurich, CH

Thoratec Switzerland GMBH appears in FDA device records in Zurich, CH. FDA records list no 510(k) clearances, no PMA records and one registered establishment. As of , FDA records show 3 recalls for this company, and at least one establishment registration is current.

FDA records held for this company

510(k) clearances
0
PMA records
0
Recall records
3
Registered establishments
1

510(k) clearance history

No 510(k) clearance records for this company in the ingested FDA data.

No 510(k) clearance records in the ingested FDA data.

Registered establishments

FDA registered establishments for Thoratec Switzerland GmbH, with the product codes listed at each
RegistrationFEIEstablishmentAddressRegistration expiresProduct codes listed
30033062483003306248Thoratec Switzerland GMBHTECHNOPARKSTRASSE 1, ZURICH CH-8005 CH2026

Registration numbers link to the FDA registration and listing record for that establishment.

FDA product codes

Recall records

Device recall records naming this company, most recent first
RecallProductReasonClassificationInitiatedStatus
Z-2215-20242nd Generation CentriMag Primary Console, REF: 201-90401, 201-90701, 201-90411, 201-90421, 201-30300, L201-90401, L201-90411, L201-90421, part of the CentriMag Circulatory Support SystemDue to a component change, the console part of a circulatory support system may not conform to the IEC 61000-4-5 standard, so if subjected to a power surge above 1.8kV and up to 2.0 kV, the console may shut down completely without alarm, pumping would cease, and the patient would have to be switched to a backup circulatory support system.Class IIOpen, Classified
Z-0221-2020CentriMag Acute Circulatory Support System Motor, U.S. Model 102956 and OUS Model 201-10002, a component of the CentriMag Extracorporeal Blood Pumping System.Reports have been received of the CentriMag Systems experiencing motor and pump issues resulting from electromagnetic interference (EMI).Class ITerminated
Z-0103-2019Thoratec CentriMag Motor, Model: 102956, a component of the Thoratec CentriMag Blood Pumping System and PediMag Blood Pumping System.Reports of circulatory support system motor stopping. Motor stop can be attributed to a break of a wire or short circuit between wires within the electrical cable where the cable exits the motor cable bend protection.Class IITerminated

Verification record

There is no verification record for this company. Most companies on this site have none: a record exists only where a company asked to be verified and MDM ran the checks, so the absence of one says nothing about the company.

Where there is a record, this is where it appears: what was checked, when it was checked, and the current status. An MDM Verified badge on another website links to a verify address on this domain that resolves to this section. A badge that does not resolve here is not ours.

What MDM checks, and what it does not

MDM coverage

MDM coverage of this company lists here, newest first. Submit news.

Is this your company?

This page is built from public FDA records. Company pages are free and factual, and stay free whether or not anyone claims them. Membership never changes editorial coverage or vetting outcomes.

Claiming lets Thoratec Switzerland GmbH edit what it says about itself and list the categories it works in. It grants no badge: claiming is not verification.

Claim this page

Data as of , FDA export datePage updated Source: openFDA, public domain