DYONICS POWER Shaver Blades, 4.5mm Long Full Radius, package of 3 Catalog Number: 7205341
Smith & Nephew, Inc. Endoscopy Division began a recall of DYONICS POWER Shaver Blades, 4.5mm Long Full Radius, package of 3 Catalog Number: 7205341 Product Usage: Smith & Nephew DYONICS Disposable Endoscopic Blades... on . The reason FDA records is "Product has insufficient weld between the shaft tube and tip, could result in the tip separating from the shaft tube during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2732-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- DYONICS POWER Shaver Blades, 4.5mm Long Full Radius, package of 3 Catalog Number: 7205341 Product Usage: Smith & Nephew DYONICS Disposable Endoscopic Blades are indicated for resection of soft and osseous tissues in both large and small articular cavities.
- Product code
- HAB Saw, Powered, And Accessories
- Reason for recall
- Product has insufficient weld between the shaft tube and tip, could result in the tip separating from the shaft tube during use.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Smith & Nephew notified customers by an "URGENT - PRODUCT RECALL 1st NOTIFICATION" letter dated May 6, 2011. The letter describes the product, problem, and actions to be taken. The letter asked customers to call 800-343-5717 (option 3) or send an email to [email protected] for a Return Authorization (RA) Number to return product, obtain credit or order replacement blades. Customers are to complete the form and return to the address provided being sure to reference the RA number. For any questions regarding this recall customers were instructed to call 508-337-3731.
- Quantity in commerce
- 555 units
- Distribution
- Worldwide Distribution -- USA (nationwide) and countries of: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, Germany, Hong Kong, Italy, Japan, Mexico, Netherlands, Turkey, and United Kingdom.
- Product codes and lots
- Lot Number: 50561144