Z-2551-2011
Naturalyte@ Sodium Bicarbonate Liquid Concentrate, Dialysate Concentrate for Hemodialysis Liquid Part number...
Fresenius Medical Care Renal Therapies Group, LLC / initiated 2011-03-23 / Terminated / root cause: manufacturing process
Fresenius Medical Care Holdings, Inc. began a recall of Naturalyte@ Sodium Bicarbonate Liquid Concentrate, Dialysate Concentrate for Hemodialysis Liquid Part number: 08-4000-LB Formulated to be used with a three... on . The reason FDA records is "Mislabeled: incorrect lot number printed on case labels". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2551-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Fresenius Medical Care Renal Therapies Group, LLC
- Product
- Naturalyte@ Sodium Bicarbonate Liquid Concentrate, Dialysate Concentrate for Hemodialysis Liquid Part number: 08-4000-LB Formulated to be used with a three stream hemodialysis machine which is calibrated for acid and bicarbonate concentrates
- Product code
- KPO Dialysate Concentrate For Hemodialysis (Liquid Or Powder)
- Reason for recall
- Mislabeled: incorrect lot number printed on case labels
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Fresenius sent an CUSTOMER NOTIFICATION LETTER letter dated April 18, 2011, to all affected customers via Certified Mail, Return Receipt Requested. Fresenius also contacted affected customers by telephone April 18, 2011, The letter identified the product, the problem and the action needed to be taken by customer. Customers were instructed to examine their stock to determine whether they have the incorrect case label and are asked to remove the correctly labeled product from the incorrectly labeled case and discard the case. Acknowledgement forms were Fax-back forms for clinics and mail-back forms (including self-addressed, stamped envelopes), for home patients. Question or concerns should be directed to Fresenius Customer Service Team at 1- 800.323.5188.
- Quantity in commerce
- 3,190 cases
- Distribution
- Nationwide Distribution--AL, CA, CT, FL, GA, NJ, NV, NY, OR, PA, and WA
- Product codes and lots
- Lot Number: 11 BMLB005 (Incorrect Lot Number: 11 BLMB005)
FIRST ARTICLE / WEEKLY