Z-2554-2011

804115, Terumo Coronary Ostia cannula, 10 Fr, 10 (25 cm) long A cardiopulmonary bypass vascular catheter...

Terumo Cardiovascular Systems Corporation / initiated 2011-05-16 / Terminated / root cause: manufacturing process

Terumo Cardiovascular Systems Corporation began a recall of 804115, Terumo Coronary Ostia cannula, 10 Fr, 10 (25 cm) long A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in... on . The reason FDA records is "Terumo has discovered several instances of adhesive build-up and or skiving of plastic (plastic flash) inside the distal tip of the Terumo Coronary Ostia Cannula. The defect was discovered during in-house inspection...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2554-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
804115, Terumo Coronary Ostia cannula, 10 Fr, 10 (25 cm) long A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.
Product code
DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Reason for recall
Terumo has discovered several instances of adhesive build-up and or skiving of plastic (plastic flash) inside the distal tip of the Terumo Coronary Ostia Cannula. The defect was discovered during in-house inspection after Terumo CVS implemented a new inspection process for the cannula. There have been no reports of the defect in clinical use.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Terumo Cardiovascular Systems Corporation sent an URGENT MEDICAL DEVICE REMOVAL notice dated May 16, 2011, to all affected customers. The notice identified the product, the problem, and the action to be taken. The letter described the problem, the potiential hazard, and the action to be taken by the consignee. Consignees were instructed to discontinue use of the product and return all product to Terumo. The letter provided a list of alternate products to be used instead of the recalled products. Consignees with questions should call 1-800-521-2818 Monday through Friday 8AM to 6PM.
Quantity in commerce
900
Distribution
Worldwide Distribution - USA (nationwide) and the countries of BELGIUM, BRAZIL, DOMINICAN REPUBLIC, UNITED ARAB EMIRATES (UAE), SINGAPORE, MALAYSIA, AND SOUTH AFRICA.
Product codes and lots
Part Number 804115 lot 0520497, 0528973, 0569610, 0606841.