Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2659-2011

Precision Ice One-Patient-Use Control Unit and Wrap System (Model # 4BB02BK). The device is intended to...

Pro Trainers' Choice Company / initiated 2011-04-18 / Terminated

Pro Trainers' Choice Company began a recall of Precision Ice One-Patient-Use Control Unit and Wrap System (Model # 4BB02BK). The device is intended to reduce edema and promote healing of soft-tissue... on . The reason FDA records is "The control panel of the Precision Ice One Patient Use Control Unit and Wrap System has a overheating issue. The control panel printed circuit boards were missing a jumper connection across resistors R28 A, B, C, and D". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2659-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Pro Trainers' Choice Company
Product
Precision Ice One-Patient-Use Control Unit and Wrap System (Model # 4BB02BK). The device is intended to reduce edema and promote healing of soft-tissue injuries and lymphedema through stimulation of the lymphatic system. The device is a therapeutic system which consists of a control unit and fluid circulating pads. The pads are mounted in wraps designed for specific parts of the body such as ankle, knee, elbow, and shoulder. The pads provide cold treatment and pressure which help promote healing. The owner's manual is labeled in part: "PRO TRAINERS' CHOICE COMPANY, INC.***www.precisionice.com***[email protected]***Mailing Address PO Box 27***Kingston, WA 98346 USA***Phone: 360-297-3902***Shipping Address***26129 Calvary Lane NE, Suite 200***Kingston, WA 98346 USA***Fax: 360-297-3903". The sticker which is placed on the back of the control board is labeled in part: "MOD#4BB02BK***SER#100***". The shipping box is labeled in part: "Pro Trainers' Choice' LLC***Model #4BB02BK***Serial #100____***". The knee wrap is labeled in part: "***Pro Trainers' Choice Co. Inc./Precision Ice One Patient Use Knee Wrap***#1BK01051KN***This device is to be used for one patient only; do not re-use!***".
Product code
ILO Pack, Hot Or Cold, Water Circulating
Reason for recall
The control panel of the Precision Ice One Patient Use Control Unit and Wrap System has a overheating issue. The control panel printed circuit boards were missing a jumper connection across resistors R28 A, B, C, and D.
FDA root cause
Other
Action
Pro Trainers' Choice sent a Mandate-Medical Device Recall on April 18, 2011 and April 20, 2011 to all affected costumers. The letter included a description of the product, affected model and lots. It instructed the costumers to discontinue using the Precision Ice, Model 4BBK02BK until the device is returned and corrected by Pro Trainers' Choice. The device should be returned to PO Box 27, Kingston, WA 98346. The customers were instructed to notify their sub-account customers of this recall if the devices were further distributed. Customers can call Pro Trainers' Choice Company at 360-297-3902 or visit the website www.precisionice.com for any questions about this recall.
Quantity in commerce
336 units
Distribution
Nationwide distribution including CA, CO, FL, IN, KS, NY, OR, TX, and WI.
Product codes and lots
The affected serial numbers range from 100116 to 100452.