Z-2610-2011

CDI TM 101 Monitor, CDI TM 101 Hematocrit/Oxygen Saturation Monitoring System, The CDI TM 101...

Terumo Cardiovascular Systems Corporation / initiated 2011-04-28 / Terminated / root cause: design

Terumo Cardiovascular Systems Corporation began a recall of CDI TM 101 Monitor, CDI TM 101 Hematocrit/Oxygen Saturation Monitoring System, The CDI TM 101 Hematocrit/Oxygen Saturation Monitoring System is intended for... on . The reason FDA records is "Terumo discovered through its own testing that the presence of electromagnetic interference (EMI) in the operating room may affect the accuracy of the values displayed by the CDI"101 Hematocrit/Oxygen Saturation...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2610-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
CDI TM 101 Monitor, CDI TM 101 Hematocrit/Oxygen Saturation Monitoring System, The CDI TM 101 Hematocrit/Oxygen Saturation Monitoring System is intended for use during cardiopulmonary bypass procedures where you need continuous monitoring of the blood's oxygen saturation, hematocrit and hemoglobin levels.
Product code
DRY Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass
Reason for recall
Terumo discovered through its own testing that the presence of electromagnetic interference (EMI) in the operating room may affect the accuracy of the values displayed by the CDI"101 Hematocrit/Oxygen Saturation Monitoring System: hematocrit (HCT), hemoglobin (Hgb) and Oxygen Saturation.
Root cause tag
design
FDA root cause
Device Design
Action
Terumo Cardiovascular Systems, Corp issued an "URGENT MEDICAL DEVICE CORRECTION" letter dated April 28, 2011, to all affected customers. The letter identifed the product, the problem and the action needed to be taken by the customer. After a solution has been verified, Terumo CVS will implement a correction for all affected units. Customers were instructed to review this Medical Device Correction notice, Assure that all users are aware of this notice. Confirm receipt of this and to fax back the Customer Response form. For further question please contact Terumo CVS Customer Service at 1-800-521-2818, Monday - Friday, 8 AM - 6 PM EST.
Quantity in commerce
622
Distribution
Worldwide Distribution -- USA ( nationwide) and the countries of Colombia, Vietnam, India, Indonesia, Pakistan, Canada, Chile, Vietnam, Chile, Hong Kong, United Arab Emirates, Singapore, Australia, Malaysia, Singapore, Taiwan, Belgium, Thailand, Vietnam and India
Product codes and lots
Lot Number 1070-1839