Z-2580-2011

MTD 4000 - Mettler Traction Decompression System, Model: ME 4000

Mettler Electronics Corp / initiated 2011-04-14 / Terminated / root cause: design

Mettler Electronics Corp began a recall of MTD 4000 - Mettler Traction Decompression System, Model: ME 4000 Product Usage: Provide traction and mobilization of skeletal structures and skeletal muscles on . The reason FDA records is "The recall was initiated because Mettler has confirmed the possible failure of internal mechanical component and software detection for potential failure regarding Electronics Traction Device - MTD4000. The firm are...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2580-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Mettler Electronics Corp
Product
MTD 4000 - Mettler Traction Decompression System, Model: ME 4000 Product Usage: Provide traction and mobilization of skeletal structures and skeletal muscles
Product code
ITH Equipment, Traction, Powered
Reason for recall
The recall was initiated because Mettler has confirmed the possible failure of internal mechanical component and software detection for potential failure regarding Electronics Traction Device - MTD4000. The firm are initiating the recall because component failure might result in patient injury. Use of the device should cease immediately.
Root cause tag
design
FDA root cause
Device Design
Action
Mettler Electronics sent an "URGENT MEDICAL DEVICE RECALL" letter dated April 14, 2011 to all affected consignees. The letter identified the product, the problem and the actions to be taken. Customers were instructed to immediately cease use of the recalled device, examine their inventory and quarantine product subject to recall. In addition, consignees were told if they have further distributed the affected product to please identify their customers and notify them at once of this product recall. The letter asks that that each includes a "Recall Return Response Form" that consignees are required to complete and return. If you have any questions please contact Mettler Electronics Corp 714-533-2221 x331, Monday through Friday, between the hours of 8:30 A.M. and 4:00 P.M. Pacific Time.
Quantity in commerce
221
Distribution
Worldwide Distribution --USA (nationwide) and countries of Taiwan, Ireland and Haiti.
Product codes and lots
Serial numbers beginning with 040, 050, 060, 070, 080, 090