Z-2740-2011

NucliSens EasyQ Enterovirus serotype 68 v1.1: Catalog number: 200300 In Vitro Diagnostic Test

bioMerieux, Inc. / initiated 2011-05-10 / Terminated / root cause: design

bioMerieux, Inc. began a recall of NucliSens EasyQ Enterovirus serotype 68 v1.1: Catalog number: 200300 In Vitro Diagnostic Test on . The reason FDA records is "The NucliSENS EasyQ Enterovirus Reagents are unable to detect Enterovirus serotype 68, in respiratory samples". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2740-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
bioMerieux, Inc.
Product
NucliSens EasyQ Enterovirus serotype 68 v1.1: Catalog number: 200300 In Vitro Diagnostic Test
Product code
OOI Real Time Nucleic Acid Amplification System
Reason for recall
The NucliSENS EasyQ Enterovirus Reagents are unable to detect Enterovirus serotype 68, in respiratory samples.
Root cause tag
design
FDA root cause
Device Design
Action
BioMerieux Vitek, Inc. sent an "Urgent Product Correction Notice" letter on/about May 10, 2011. The letter included; a description of the product, problem, possible impact to the patient and recommendations. Recommendations included instructing customers to discontinue the use of the kit for the detection of Enterovirus 68 and Rhinovirus 87 serotypes and noted the corrected IFU will be available the first week of April 2011. An Acknowledgement form was attached to be returned via fax. For additional questions please contact your local BioMerieux Clinical Customer representative at (800) 682-2666, option 3, option 2.
Quantity in commerce
42 kits
Distribution
(USA) Nationwide distribution including the states of MD, MI, MN, MT, NY and PA
Product codes and lots
Catalog number: 200300, Lot number: 10042001, 9081001, 9112403, 9030902 and 9030901.