Z-3126-2011

Access Immunoassay Systems Part Number: 81600 Access 2 Immunoassay Systems Part Number: 81600N Access...

Siemens Healthcare Diagnostics Inc. / initiated 2011-04-15 / Terminated

Beckman Coulter Inc. began a recall of Access Immunoassay Systems Part Number: 81600 Access 2 Immunoassay Systems Part Number: 81600N Access Immunoassay Systems LXi 725 Part Number: 386200... on . The reason FDA records is "The recall was initiated because certain Access immunoassays can produce inaccurate results when performed within the published UniCel DxI Immunoassay Systems* room temperature operating specification of 18 degrees C to...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-3126-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Access Immunoassay Systems Part Number: 81600 Access 2 Immunoassay Systems Part Number: 81600N Access Immunoassay Systems LXi 725 Part Number: 386200 Subsequent Product Codes: JGS The Access/DxI Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
The recall was initiated because certain Access immunoassays can produce inaccurate results when performed within the published UniCel DxI Immunoassay Systems* room temperature operating specification of 18 degrees C to 32 degrees C (64.4 degrees F to 89.6 degrees F). An increase in room temperature causes assay results for some assays to decrease, while results for other assays increase with an i
FDA root cause
Other
Action
Beckman Coulter sent an "URGENT FIELD SAFETY NOTICE" letter dated April 18, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter includes information on the new room temperature operating range for all UniCel DxI Immunoassay Systems and Access Immunoassay Systems during system operation. Customers were instructed to review their historical room temperature records to determine if their Access Immunoassay System and UniCel DxI Immunoassay System were operating outside the revised temperature range. An attached Response Form was included for customers to complete and return. Please contact Beckman Coulter Customer Support Center for questions regarding this notice at 1-800-854-3633.
Quantity in commerce
11,266 units
Distribution
Worldwide Distribution--USA (nationwide) and Puerto Rico and the countries of Algeria, Angola, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Czech Republic, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Korea, Republic of, Kuwait, Lebanon, Liechtenstein, Malaysia, Mayette, Mexico, Morocco, Nambia, Netherlands, New Zealand, Norway, Philippines, Poland,, Romania, Russian Federation, Singapore, Slovakia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, United States, Venezuela and Vietnam.
Product codes and lots
All Serial Numbers