Z-2600-2011
Prism Laser 7642 Laser System
American Stress Technologies, Inc. / initiated 2011-04-27 / Terminated / root cause: labeling and UDI
American Stress Technologies, Inc. began a recall of Prism Laser 7642 Laser System on . The reason FDA records is "It was discovered The Prism 7642 was not certified at the time of delivery. The product lacks the required labeling and the product was sold and delivered without the submission of the Initial Product Report. The...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2600-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- American Stress Technologies, Inc.
- Product
- Prism Laser 7642 Laser System
- Product code
- REO Analytical Measuring And Detection Laser Product
- Reason for recall
- It was discovered The Prism 7642 was not certified at the time of delivery. The product lacks the required labeling and the product was sold and delivered without the submission of the Initial Product Report. The product lacks User In formation
- Root cause tag
- labeling and UDI
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- American Stress Technologies sent a customer notification letter dated April 1, 2011, informing them of the action they have taken to correct this issue. (1) ASTI conducted laser safety tests and checks on May 3,2011 to verify the Prism was compliant with FDA regulations. This was confirmed in a June 17, 2011, electronic mail message. (2) ASTI placed the required labels described in the initial product report on the Prism on May 3, 2011. (3) ASTI delivered updated user manuals to UNO on May 3, 2011, that describe potential laser hazards and the required labeling in the same way as described in the initial product report. (4) ASTI completed the above corrective actions at no cost to UNO as confirmed in a June 20, 2011, electronic mail message. If you have any further questions please call (724) 410-1030
- Quantity in commerce
- 1
- Distribution
- Nationwide Distribution
- Product codes and lots
- S/N 7642/2009