Kinamed SuperCable Grip and Plate System Driver, 3.5mm Hex, Catalog Number: 35-860-2060 The driver is...

Kinamed Inc / initiated 2010-08-18 / Terminated / root cause: manufacturing process

Kinamed Inc began a recall of Kinamed SuperCable Grip and Plate System Driver, 3.5mm Hex, Catalog Number: 35-860-2060 The driver is intended to be used in conjunction with compression and... on . The reason FDA records is "The recall was initiated due to inadequate mechanical strength of these screwdrivers' hexagonal tip feature, causing it to strip during final tightening". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2588-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Kinamed Inc
Product
Kinamed SuperCable Grip and Plate System Driver, 3.5mm Hex, Catalog Number: 35-860-2060 The driver is intended to be used in conjunction with compression and locking bone screws for fixation of long bone fractures when Kinamed Trochanteric Grips & Cable-Plates are used. Subsequent Product Codes: HWC, JDQ.
Product code
KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
Reason for recall
The recall was initiated due to inadequate mechanical strength of these screwdrivers' hexagonal tip feature, causing it to strip during final tightening.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm, Kinamed Inc., sent a recall letter dated August 18, 2010 to consignees/customers via FedEx. The letter described the product, problem and actions to be taken. The customers were instructed to Immediately examine their stock and return any Kinamed SuperCable Screwdriver, 3.5mm Hex bearing these lot numbers, along with the completed attached Effectiveness Check Form and shipment under RMA #74567 to the attention of: "Director of Quality Assurance" Kinamed, Inc. 820 Flynn Road Carnarillo, CA 93012-8701 Note: The Kinamed sales agent shipped and had the product replaced for the customers. If you have any questions, please contact (303)-681-2208, (805) 384-2748 or (800) 827-5775
Quantity in commerce
8 units
Distribution
Worldwide distribution: USA (nationwide) state of: CO and country of: Columbia.
Product codes and lots
Lot Numbers K12899 (2pcs) & K12985 (3pcs)