Z-2637-2011

MEVATRONKD, camera imaging system, Part Number: 8319857; Product is manufactured and distributed by SIEMENS...

Siemens Medical Solutions USA, Inc. / initiated 2010-11-04 / Terminated

Siemens Medical Solutions USA, Inc - Radiation Oncology began a recall of MEVATRONKD, camera imaging system, Part Number: 8319857; Product is manufactured and distributed by SIEMENS MEDICAL SOLUTIONS USA, INC. Concord, CA; and... on . The reason FDA records is "The firm has become aware of a potential issue that may result in a service engineer receiving an electrical shock should the connector cable housing become unfastened from the chassis". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2637-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
MEVATRONKD, camera imaging system, Part Number: 8319857; Product is manufactured and distributed by SIEMENS MEDICAL SOLUTIONS USA, INC. Concord, CA; and SIEMENS AG, MEDICAL SOLUTIONS Roentgenstrasse, Kemnath, GERMANY, and SIEMENS AG, MEDICAL SOLUTIONS Henkestrasse, Erlangen, GERMANY The intended use of the SIEMENS branded linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
The firm has become aware of a potential issue that may result in a service engineer receiving an electrical shock should the connector cable housing become unfastened from the chassis.
FDA root cause
Other
Action
The firm, SIEMENS, sent a "Customer Information" letter to its customers on November 4, 2010. The letter describes the product, problem and actions to be taken. The customers were instructed to include this letter in their BEAMVIEW TI System Owner Manual chapter "Safety Advisory Letters" where it should remain. Siemens has developed an updated kit to install an additional grounding cable on the camera's central control unit to eliminate the problem. The updated kit was delivered with the Customer Information letter. If you have any questions, contact the Service Manager- Regulatory Affairs at 925-246-8407.
Distribution
Worldwide distribution:
Product codes and lots
All units of the part number: 8319857