Jet-Ventilator Catheter for Adults, 14G, STERILE, NO LATEX, REF 30-02-914-1, VBM Medizintechnik GmbH...
VBM Medical Inc. began a recall of Jet-Ventilator Catheter for Adults, 14G, STERILE, NO LATEX, REF 30-02-914-1, VBM Medizintechnik GmbH, Einsteinstrasse 1 7212 Sulz a.N. Germany Allows a safe... on . The reason FDA records is "Product labeled with incorrectly as being appropriate for use in adults when the intended population for use of the product is children and infants". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2675-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- VBM Medical Inc.
- Product
- Jet-Ventilator Catheter for Adults, 14G, STERILE, NO LATEX, REF 30-02-914-1, VBM Medizintechnik GmbH, Einsteinstrasse 1 7212 Sulz a.N. Germany Allows a safe and atraumatic transtracheal access for oxygenation/ventilation with a manual jet ventilator or an automatic high frequency jet ventilator.
- Product code
- BTO Tube, Tracheostomy (W/Wo Connector)
- Reason for recall
- Product labeled with incorrectly as being appropriate for use in adults when the intended population for use of the product is children and infants.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- VBM Medical Inc sent an "URGENT FIELD SAFETY NOTICE" dated May 9, 2011, to all affected customers. The letter identified the product, the problem, and the action needed to be taken by the customer. The reference and lot numbers, details of the error, along with the incorrect and correct box label, and instructions to the customers to discard the box with the incorrect label or send the complete product back to VBM Medical Inc. for re-packaging. For any further or detailed questions, please contact VBM Medical Inc. at (317) 776-1800. On 7/12/11 upon receiving the recall classification letter, the firm sent out an additional letter to its direct accounts informing them of the recall's classification and reiterating the instructions listed above.
- Quantity in commerce
- 388
- Distribution
- Worldwide Distribution -- USA ( nationwide ) including the country Montreal QC
- Product codes and lots
- Lot numbers 28327, 30067, 31065, 32642 and article number ACU 1060.2