syngo Dynamics Picture Archiving Communication System The software is a picture archiving and communication...
Siemens Medical Solutions, USA, Inc began a recall of syngo Dynamics Picture Archiving Communication System The software is a picture archiving and communication system intended for acceptance, transfer, display... on . The reason FDA records is "For systems upgraded to syngo Dynamics version 9.0, if a legacy report generated by version 3.x or earlier is opened in the version 9.0 portal, the legacy report may be deleted and replaced with a new report from the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2762-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- syngo Dynamics Picture Archiving Communication System The software is a picture archiving and communication system intended for acceptance, transfer, display, storage, archive, and manipulation of digital medical images including quantification and report generation.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- For systems upgraded to syngo Dynamics version 9.0, if a legacy report generated by version 3.x or earlier is opened in the version 9.0 portal, the legacy report may be deleted and replaced with a new report from the version 9.0 reporting system or a blank report.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens Medical Solutions, USA, Inc sent a letter entitled "Customer Safety Advisory Notice" to all direct accounts on May 10, 2011. The letter stated the name and version of the software, problem noted, a work-around to be implemented until a service patch is created to fix the problem, and that the letter should be placed in the software's operator's manual to be seen by all who utilize the software.
- Quantity in commerce
- 38
- Distribution
- Nationwide Distribution
- Product codes and lots
- Serial numbers 10091602 and 10091604
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.