Pressure Monitoring Kit, Catalog 16066100, Terumo Cardiovascular Systems Ann Arbor MI. The pressure...
Terumo Cardiovascular Systems Corporation began a recall of Pressure Monitoring Kit, Catalog 16066100, Terumo Cardiovascular Systems Ann Arbor MI. The pressure monitoring kit is indicated for sterile single use with the... on . The reason FDA records is "The firm has recieved complaints regarding 35 pressure monitoring kits in which the membrane in the kit separated form the pressure dome housing. The firm determined that the defect was caused by a change in the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2613-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Pressure Monitoring Kit, Catalog 16066100, Terumo Cardiovascular Systems Ann Arbor MI. The pressure monitoring kit is indicated for sterile single use with the reusable pressure transducer.
- Product code
- DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Reason for recall
- The firm has recieved complaints regarding 35 pressure monitoring kits in which the membrane in the kit separated form the pressure dome housing. The firm determined that the defect was caused by a change in the pressure dome's manufacturing process.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Terumo Cardiovascular Systems sent an URGENT MEDICAL DEVICE REMOVAL notice dated May 3, 2011, to all affected customers. The letter identified the product, the problem and the action needed to be taken by the customer. Customers were sent 10 replacement pressure monitoring kits with the removal notice as immediate replacements. The firm will send additional replacements to customers that have more than 10 units of affected product as noted on the Customers' returned reponse form. Customers that received the custom packs were provided stickers to identify the recalled packs. The customers were instructed not to open the packs to remove the affected pressure monitoring kit prior to clinical use, but to replace the pressure domes at the time of use. For more information please contact Terumo Cardiovascular Customer Service at 800- 521-2818.
- Quantity in commerce
- 19495 units
- Distribution
- Worldwide Distribution -- USA (nationwide )including the states AR, CA, FL, GA, MO, MT, NC, ND, OK, OR, PA, TN, TX, UT, VA, WA and WY, and the countries of Argentina, Belgium, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, Guatamala, Honduras, Hong Kong, Indonesia, Kuwait, Malaysia, Mexico, Pakistan, Panama, Philippines, Singapore, South Korea, Thailand, United Arab Emirates, Uruguay and Vietnam
- Product codes and lots
- 0511829, 0519808, 0522673, 0537066, 0538520, 0544571, 0548624, 0552118, 0555878, 0557544, 0563808, 0565139, 0571582, 0575070, 0578021, 0578529, 0579318, 0580173, 0582267, 0584399, 0587165, 0589587, 0592742, 0593799, 0595311, 0595312, 0604098.