Pinnacle Pelvic Floor Repair Kit--Anterior/Apical STERILE, and Pinnacle Pelvic Floor Repair Kit--Posterior...
Boston Scientific Corporation began a recall of Pinnacle Pelvic Floor Repair Kit--Anterior/Apical STERILE, and Pinnacle Pelvic Floor Repair Kit--Posterior STERILE. Boston Scientific, Corp. Indicated for... on . The reason FDA records is "The device may exhibit low tensile strength between the needle and suture and lead to needle detachment during mesh leg placement". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2931-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Pinnacle Pelvic Floor Repair Kit--Anterior/Apical STERILE, and Pinnacle Pelvic Floor Repair Kit--Posterior STERILE. Boston Scientific, Corp. Indicated for tissue reinforcement and stabilization of fascial structures of the pelvic floor for vaginal wall prolapse, where surgical treatment is intended, either as mechanical support or bridging material for the fascial defect.
- Product code
- FTL Mesh, Surgical, Polymeric
- Reason for recall
- The device may exhibit low tensile strength between the needle and suture and lead to needle detachment during mesh leg placement.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Boston Scientific Corporation sent on May 10, 2011 an "URGENT MEDICAL DEVICE RECALL-IMMEDIATE ACTION REQUIRED" letter to all affected customers. The letter included; reason for recall, instructions for the quarantine, discontinued use, and return of the devices, and information for Medwatch reporting. The letter also includes a form that is to be faxed to the firm. For information call (508) 683-4678.
- Quantity in commerce
- 540
- Distribution
- Worldwide Distribution; (USA) Nationwide distribution, and the countries of Puerto Rico, France, Great Britain, Ireland, Italy, Spain, Sweden, Canada, and Australia.
- Product codes and lots
- Catalog/UPC: M0068317100, Lot: 1ML0072001. Catalog/UPC M0068317050, Lot: 1ML0070801.