Z-2581-2011
Cuffable Blood Pressure Cuffs; Manufactured for Vital Signs, Inc., a GE Healthcare Company 20 Campus Road...
GE Healthcare / initiated 2011-04-26 / Terminated / root cause: design
GE Healthcare, LLC began a recall of Cuffable Blood Pressure Cuffs; Manufactured for Vital Signs, Inc., a GE Healthcare Company 20 Campus Road, Totowa, NJ 07512 Made in Mexico. The cuff is used to... on . The reason FDA records is "BP cuffs may not properly inflate due to a leak". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2581-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Cuffable Blood Pressure Cuffs; Manufactured for Vital Signs, Inc., a GE Healthcare Company 20 Campus Road, Totowa, NJ 07512 Made in Mexico. The cuff is used to determine a subject's blood pressure.
- Product code
- DXQ Blood Pressure Cuff
- Reason for recall
- BP cuffs may not properly inflate due to a leak
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- GE Healthcare sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated April 26, 2011 to all affected customers. The letter identified the product, the problem and the action to be taken. Customers were instructed to discontinue use of the product, isolate all affected products, return only unused products and distribute this letter to any other healthcare institutions that are potentially affected by this issue. The letter asks that each customer complete and return the enclosed confirmation form via fax to 800-535-7923, in order to confirm receipt of the notification letter. For any questions, contact your local Vital Signs Account Manager or the International Customer Service at 1-800-932-0760.
- Quantity in commerce
- 2,684,996 units
- Distribution
- Worldwide Distribution -- USA (nationwide) including Puerto Rico and countries of: Argentina, Australia, Belgium, Canada, Chile, Ecuador, Finland, Germany, Hong Kong, Israel, Italy, Latvia, Lebanon, Nicaragua, Panama, Peru, Saudi Arabia, Slovenia, UK, Uruguay and Venezuela
- Product codes and lots
- 510(k) K911213 Medical Device Listing Number D110124 for year 2009- Lot #2689- Lot #3659 For year 2010 - Lot #0010-Lot #2530