Z-2626-2011

BACTEC" FX System Software, catalog #441398 and #441750 The BD BACTEC" FX instrument is designed for the...

Becton Dickinson & Co / initiated 2010-08-10 / Terminated / root cause: software

Becton Dickinson & Co. began a recall of BACTEC" FX System Software, catalog #441398 and #441750 The BD BACTEC" FX instrument is designed for the rapid detection of bacteria and fungi in clinical... on . The reason FDA records is "In vitro diagnostic instrumentation may exhibit random operational errors that could delay effective patient treatment". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2626-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
BACTEC" FX System Software, catalog #441398 and #441750 The BD BACTEC" FX instrument is designed for the rapid detection of bacteria and fungi in clinical specimens. Samples are drawn from patients and injected directly into BACTEC culture vials, which are placed into the instrument for incubation and testing.
Product code
MDB System, Blood Culturing
Reason for recall
In vitro diagnostic instrumentation may exhibit random operational errors that could delay effective patient treatment.
Root cause tag
software
FDA root cause
Software design
Action
BD Diagnostics Systems sent an "URGENT PRODUCT RECALL" letter dated August 2010 to all affected customers. The letter describes the product, problem and action to be taken. The letter described the type of error messages the customers may experience. All affected units will be scheduled for on site software update as well as a disk drive replacement. A BD field service engineer will contact the customers to schedule the on site visit. Customers experiencing issues prior to the scheduled visit or have questions should contact the BD Diagnostics Service Department at 800-638-8663.
Quantity in commerce
907 units
Distribution
Worldwide Distribution--USA (nationwide) including countries of Australia, Belgium, Canada, China, Colombia, New Zealand, Hong Kong, Japan, Korea, Philippines, Singapore, Taiwan, and Thailand.
Product codes and lots
Catalog # 441398-Serial numbers:V100E, V101F, V200B, V210A, V220B, V221A, and V230B  Catalog # 441750-Serial number: V100A