Z-2786-2011

Siemens MAGNETOM Systems Symphony, Sonata and Symphony a TIM System Nuclear magnetic resonance imaging system

Siemens Medical Solutions USA, Inc. / initiated 2011-03-31 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA, Inc began a recall of Siemens MAGNETOM Systems Symphony, Sonata and Symphony a TIM System Nuclear magnetic resonance imaging system on . The reason FDA records is "Firm has discovered during product monitoring that on a few systems the gradient cable connections did not meet their specifications after installation". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2786-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens MAGNETOM Systems Symphony, Sonata and Symphony a TIM System Nuclear magnetic resonance imaging system
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
Firm has discovered during product monitoring that on a few systems the gradient cable connections did not meet their specifications after installation.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Siemens initated their recall of the device on March 31, 2001, by sending a Customer Safety Advisory Notice/Letter to affected consignees. The letter identified the product, the problem, and the action to be taken by the customer. Customers were notified that detectors will be installed in appropriate locations on the system, which will switch off the system's power electronics in the event of a malfunction in the power supply of the gradient coils, preventing damage to the system and its environment.
Quantity in commerce
440
Distribution
Nationwide Distribution including AL, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, and WI and Puerto Rico.
Product codes and lots
Model numbers 4760265, 7104594, 7104719, 7104735, 7106425 and 7106557.  Serial numbers 21604, 21235, 14151, 13314, 14312, 22499, 13438, 22301, 22211, 22311, 14247, 22127, 14436, 14228, 22518, 22165, 22200, 14306, 14426, 13044, 14138, 22438, 14023, 21204, 22344, 14196, 22195, 22276, 22355, 14424, 14369, 14393, 21279, 22491, 14114, 22309, 14344, 14101, 22386, 22523, 14373, 22609, 22305, 22090, 22561, 14394, 14292, 22580, 22524, 22568, 22573, 14144, 22173, 22270, 22511, 14009, 14087, 14175, 22170, 22604, 22115, 14024, 22114, 14383, 22352, 22121, 14097, 14486, 14155, 14234, 21250, 22587, 14204, 22243, 22408, 14060, 13121, 22621, 14323, 22557, 22250, 22029, 21936, 22356, 22079, 22342, 14405, 14338, 22126, 14379, 14015, 22618, 14360, 22463, 22307, 14237, 14206, 14276, 14278, 22418, 22649, 21928, 22643, 14343, 14188, 14376, 14409, 22153, 22642, 22312, 22084, 14387, 22444, 14448, 14314, 22097, 14429, 14072, 21200, 14390, 14348, 14229, 21608, 22543, 21253, 14341, 13108, 22221, 14488, 14066, 22647, 22193, 13523, 14389, 22610, 22321, 14299, 14484, 22316, 14447, 21108, 22298, 14158, 22591, 14231, 14487, 22623, 14385, 22302, 22196, 14201, 14225, 22169, 14368, 22277, 14252, 14319, 22075, 14328, 14265, 13413, 13345, 22237, 21268, 22629, 14444, 14450, 14431, 21926, 22324, 22225, 22241, 22265, 14076, 22483, 14221, 14220, 14346, 22330, 22108, 22650, 22617, 22403, 22134, 22212, 22047, 21174, 22319, 14148, 22608, 22286, 22357, 13135, 21105, 21155, 21201, 22092, 14363, 22145, 14415, 21142, 22322, 14371, 14417, 14438, 22046, 22534, 22441, 22010, 22080, 22435, 14182, 22021, 14198, 21205, 14219, 14270, 14412, 14437, 22536, 22248, 22471, 22503, 22506, 14119, 22242, 14258, 14474, 14452, 22306, 14357, 21611, 14010, 22616, 21262, 22216, 22171, 21119, 22576, 22174, 22407, 14381, 21186, 22162, 22103, 22228, 14386, 14161, 22605, 22346, 22406, 14443, 22565, 22639, 22105, 14380, 22638, 21189, 14044, 14413, 14186, 22215, 14432, 22582, 14423, 22130, 22497, 22024, 22490, 14105, 21906, 14356, 22164, 22093, 22345, 14321, 22636, 14398, 14355, 22607, 13255, 22189, 22190, 22478, 22400, 22579, 22218, 22273, 14041, 14126, 22073, 22451, 14184, 14256, 14288, 22588, 22091, 14232, 14401, 21171, 13110, 22519, 14043, 22271, 22051, 22261, 14453, 22255, 22031, 22520, 22455, 21289, 21238, 22011, 14137, 22551, 14345, 22129, 22207, 14235, 22063, 21249, 22372, 22083, 22402, 13325, 22358, 22572, 22416, 14449, 14180, 21914, 14193, 22113, 22184, 22569, 22433, 22493, 22172, 22606, 14092, 14272, 14427, 14085, 14399, 22622, 22624, 14038, 22266, 22060, 14223, 21202, 22398, 22461, 22244, 21131, 14035, 13475, 22095, 14322, 14433, 22068, 14115, 13415, 14411, 22036, 22050, 14093, 14142, 14313, 14324, 14211, 14261, 22646, 14349, 22429, 22182, 22574, 22625, 22361, 14460, 14146, 21260, 21612, 14160, 22118, 21904, 14215, 14013, 14251, 14397, 22071, 14128, 21107, 21264, 14109, 14430, 21239, 21236, 14177, 14055, 14226, 21232, 22401, 13114, 22272, 22530, 14472, 22596, 21138, 21299, 14189, 14230, 14245, 22464, 21215, 22282, 14036, 14048, 22339, 22602, 22535, 22510, 14334, 22074, 21251, 14340, 22648, 14238, 22615, 14120, 22104, 14434, 22030, 21165, 14362, 22175, 22550, 21248, 14051.