BBL" OXACILLIN SCREEN AGAR, catalog #221952, box of 10 plates. Oxacillin Screen Agar (originally named MRSA...
Becton Dickinson & Co. began a recall of BBL" OXACILLIN SCREEN AGAR, catalog #221952, box of 10 plates. Oxacillin Screen Agar (originally named MRSA Screen Agar) was developed for the detection of... on . The reason FDA records is "Oxacillin Screen Agar may not have been produced according to specification and could result in false antibiotic susceptibility test results in patient specimens". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2608-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BBL" OXACILLIN SCREEN AGAR, catalog #221952, box of 10 plates. Oxacillin Screen Agar (originally named MRSA Screen Agar) was developed for the detection of methicillin-resistant Staphylococcus aureus (MRSA). These strains are resistant to penicillinase-resistant penicillins (PRPs), such as methicillin, oxacillin and nafcillin. Since the method to detect MRSA uses the same inoculum as the Bauer-Kirby antimicrobial disc susceptibility test procedure, the oxacillin screen test may be conveniently performed on isolates at the same time as routine susceptibility testing.
- Product code
- JTZ Culture Media, Antimicrobial Susceptibility Test, Mueller Hinton Agar/Broth
- Reason for recall
- Oxacillin Screen Agar may not have been produced according to specification and could result in false antibiotic susceptibility test results in patient specimens.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- BD Diagnostic Systems sent an "URGENT PRODUCT RECALL" letter dated April 2011 to all affected customers. The letter identifies the product, problem and actions to be taken by the customers. The letter instructed customers to discontinue use of the affected lot number and discard any remaining packages. Customers were requested to complete and return an attached response form via fax at 410-316-4258. The letter states that all discarded product will be replaced. Questions concerning the recall process were directed to BD Customer Service Department at 1-800-675-0908.
- Quantity in commerce
- 5440 plates
- Distribution
- Worldwide Distribution--USA (nationwide) and the countries of Canada, Columbia, Jamaica, South Korea, and Taiwan.
- Product codes and lots
- Lot 1034119 Exp. 2011/05/03