Z-2674-2011

Spectra Optia Apheresis System, REF: 61000, CaridianBCT, Lakewood, CO 80215. The Spectra Optia Apheresis...

Terumo BCT, Ltd / initiated 2010-12-09 / Terminated / root cause: software

Caridian BCT, Incorporated began a recall of Spectra Optia Apheresis System, REF: 61000, CaridianBCT, Lakewood, CO 80215. The Spectra Optia Apheresis System, a blood component separator, is intended for... on . The reason FDA records is "An additional air detection system is being added to all existing systems to detect air in the system's return line". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2674-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo BCT, Ltd
Product
Spectra Optia Apheresis System, REF: 61000, CaridianBCT, Lakewood, CO 80215. The Spectra Optia Apheresis System, a blood component separator, is intended for use in therapeutic apheresis applications. The system may be used to perform Therapeutic Plasma Exchange, Mononuclear Cell Collection, and Red Blood Cell Exchange.
Product code
LKN Separator, Automated, Blood Cell And Plasma, Therapeutic
Reason for recall
An additional air detection system is being added to all existing systems to detect air in the system's return line.
Root cause tag
software
FDA root cause
Software design
Action
CaridianBCT sent a "MEDICAL DEVICE RECALL" letter dated December 2010 to all customers. The letter describes the product, problem, and actions to be taken by the customers. Customers were advised to continue using their systems as long as they followed the instructions in the Operator's Manual. An Acknowledgement of Receipt form was included with the letter to be completed and faxed to 1-303-876-9277 or e-mailed to [email protected]. A CaridianBCT representative will schedule a visit to facilities to upgrade their systems within the next 6 months. Further information is available by contacting the CaridianBCT Support Center at 877-339-4228 or 303-231-4357.
Quantity in commerce
594 units
Distribution
Worldwide Distribution--USA (nationwide) and countries of Australia, Belgium, Canada, China, Hong Kong, Japan, Malaysia, Mexico, Pakistan, Saudi Arabia, Singapore, South Korea, Taiwan, Thailand, and Vietnam.
Product codes and lots
Serial Numbers: 1P00100 thru 1P00719