***REF 66800694***Qty 10***smith&nephew***RENASYS Port Foam Dressing Kit Port***STERILE***Rx Only.***Made in...
Smith & Nephew Inc. began a recall of ***REF 66800694***Qty 10***smith&nephew***RENASYS Port Foam Dressing Kit Port***STERILE***Rx Only.***Made in Mexico for/Fabrique au Mexique pour: Smith &... on . The reason FDA records is "On 12/14/2010, Smith & Nephew, Inc. Wound Management Division in St. Petersburg, FL recalled their RENASYS PORT product code number 66800694 due to small holes in the primary pouch of some of the packaged units of which...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0875-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- ***REF 66800694***Qty 10***smith&nephew***RENASYS Port Foam Dressing Kit Port***STERILE***Rx Only.***Made in Mexico for/Fabrique au Mexique pour: Smith & Nephew, Inc., St. Petersburg, FL 33716***Smith & Nephew Medical Limited Hull HU3 2BN England***Lot 123456. Product Usage: The product is a suction device used as an accessory to the foam dressing kits, which is used in conjunction with Smith & Nephew Negative Pressure Wound Therapy (NPWT) systems to deliver negative pressure to a wound.
- Product code
- OMP Negative Pressure Wound Therapy Powered Suction Pump
- Reason for recall
- On 12/14/2010, Smith & Nephew, Inc. Wound Management Division in St. Petersburg, FL recalled their RENASYS PORT product code number 66800694 due to small holes in the primary pouch of some of the packaged units of which constituted a breach of the sterile barrier.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Smith & Nephew, Inc., Wound Management sent an URGENT DEVICE RECALL letter dated December 14, 2010, to all affected customers. The letter identified the product, problem, and actions to be taken. Customers were instructed to immediately identify and isolate the recalled lots in their inventory, complete and fax the response form, and return any affected products to APL Logistics, Inc. in Suwanee, GA. For questions in general, please contact the Customer Care Center at 1-800-876-1261.
- Quantity in commerce
- 71329 units
- Distribution
- Worldwide Distribution - USA (nationwide) including the states of: AZ, CA, GA, MA, MI, MO, MS,.NC, NY,OR, TN, and WI. and the countries of: Australia, Canada, China, England, Germany, Ireland and Singapore.
- Product codes and lots
- Product Code Number: 66800694 Lot numbers: 50512962, 50512969, 50512974, 50513087, 50514672, 50513073, 50513074, 50517647, 50516110, 50539644, 50539645, 50539646, 50539647, 50555348, 50555349, 50555351, 50539648, 50546339, 50546340, 50548244, 50548245, 50548246, 50548247 and 50548248.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.