Z-2721-2011

ECMO Heater Model 333W, Cardiopulmonary bypass temperature controllers. This device is intended to provide...

Gentherm Medical, LLC / initiated 2011-05-02 / Terminated / root cause: design

Cincinnati Sub-Zero Products Inc began a recall of ECMO Heater Model 333W, Cardiopulmonary bypass temperature controllers. This device is intended to provide temperature-controlled warm water flow to a blood... on . The reason FDA records is "The probe jack contacts (terminals) on certain of the firm's ECMO Heater devices are located too close to the solid state relay contacts (terminals) which can lead to the terminals coming into contact with each other...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2721-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Gentherm Medical, LLC
Product
ECMO Heater Model 333W, Cardiopulmonary bypass temperature controllers. This device is intended to provide temperature-controlled warm water flow to a blood heat exchanger in order to maintain a patient's blood temperature. It can be used during extracorporeal circulation procedures. It is intended to use with commercially available heat exchangers.
Product code
DWJ System, Thermal Regulating
Reason for recall
The probe jack contacts (terminals) on certain of the firm's ECMO Heater devices are located too close to the solid state relay contacts (terminals) which can lead to the terminals coming into contact with each other. If this situation occurs, the service technician, user, and/or patient could be exposed to electric shock during servicing and/or use.
Root cause tag
design
FDA root cause
Device Design
Action
Cincinnati Sub-Zero Products Inc issued an "Urgeant Medical Device Field Action" letter on May 2, 2011 to all affected customers. The letter includes; description of product and problem, affected lots and instructs customers to discontinue its use. The product is to be returned to the Service Department for an update and the customer is instructed to complete and mail Field Action Form. For additional contact the firm at 1-800-989-7373 or (513) 772-8810.
Quantity in commerce
212 units - U. S. distribution & 44 units foreign distribution
Distribution
Worldwide Distribution and countries of The Netherlands, Kuwait, India, Saudi Arabia, Taiwan and Australia. (USA) Nationwide including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IL, IN, LA, MA, MD, MI, MIN, MO, NC, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, WA, and WI.
Product codes and lots
Model: 333W Part Number: 86135 (115V), 86136 (230V) and 86138 (240V) Serial Number: 942-E2979 through 031-E5150