VITEK 2 System, VITEK 2 60 110V (Catalog #27225) and VITEK 2 60 220V (Catalog #27226). VITEK 2 is an...

bioMerieux, Inc. / initiated 2011-06-02 / Terminated / root cause: design

Biomerieux Inc began a recall of VITEK 2 System, VITEK 2 60 110V (Catalog #27225) and VITEK 2 60 220V (Catalog #27226). VITEK 2 is an automated system consisting of instruments, software, and... on . The reason FDA records is "Cards may not load into the incubator correctly causing them to be loaded in the wrong order which could result in the results being linked to the wrong accession number and reporting results erroneously". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2767-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
bioMerieux, Inc.
Product
VITEK 2 System, VITEK 2 60 110V (Catalog #27225) and VITEK 2 60 220V (Catalog #27226). VITEK 2 is an automated system consisting of instruments, software, and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The system is intended for use to perform in vitro diagnostic organism identification and antimicrobial susceptibility testing in human applications.
Product code
LTW Susceptibility Test Cards, Antimicrobial
Reason for recall
Cards may not load into the incubator correctly causing them to be loaded in the wrong order which could result in the results being linked to the wrong accession number and reporting results erroneously.
Root cause tag
design
FDA root cause
Device Design
Action
Biomerieux, Inc. sent an "URGENT PRODUCT CORRECTION NOTICE" dated June 2, 2011 to all affected customers. The letter describes the products, problem, and actions to be taken by the customers. The letter explains how to recognize the issue and instructs customers to take certain steps to resolve the problem. Additionally, an Acknowledgement Form is attached to the letter for customers to complete and return via fax to 919-314-4539. A firmware update will be applied by Biomerieux, Inc. technical support representatives. Questions regarding this recall are directed to the Customer Service Department at 1-800-682-2666, option 3.
Quantity in commerce
627 instruments
Distribution
Worldwide Distribution-- USA (nationwide) and the countries of Austria, Australia, Belgium, Brazil, Canada, China, Colombia, Germany, France, India, Israel, Italy, Japan, Korea, Mexico, The Netherlands, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, United Arab Emirate, and United Kingdom.
Product codes and lots
VTK23119 to VTK 23745