VITEK 2 System, VITEK 2 60 110V (Catalog #27225) and VITEK 2 60 220V (Catalog #27226). VITEK 2 is an...
Biomerieux Inc began a recall of VITEK 2 System, VITEK 2 60 110V (Catalog #27225) and VITEK 2 60 220V (Catalog #27226). VITEK 2 is an automated system consisting of instruments, software, and... on . The reason FDA records is "Cards may not load into the incubator correctly causing them to be loaded in the wrong order which could result in the results being linked to the wrong accession number and reporting results erroneously". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2767-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- bioMerieux, Inc.
- Product
- VITEK 2 System, VITEK 2 60 110V (Catalog #27225) and VITEK 2 60 220V (Catalog #27226). VITEK 2 is an automated system consisting of instruments, software, and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The system is intended for use to perform in vitro diagnostic organism identification and antimicrobial susceptibility testing in human applications.
- Product code
- LTW Susceptibility Test Cards, Antimicrobial
- Reason for recall
- Cards may not load into the incubator correctly causing them to be loaded in the wrong order which could result in the results being linked to the wrong accession number and reporting results erroneously.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Biomerieux, Inc. sent an "URGENT PRODUCT CORRECTION NOTICE" dated June 2, 2011 to all affected customers. The letter describes the products, problem, and actions to be taken by the customers. The letter explains how to recognize the issue and instructs customers to take certain steps to resolve the problem. Additionally, an Acknowledgement Form is attached to the letter for customers to complete and return via fax to 919-314-4539. A firmware update will be applied by Biomerieux, Inc. technical support representatives. Questions regarding this recall are directed to the Customer Service Department at 1-800-682-2666, option 3.
- Quantity in commerce
- 627 instruments
- Distribution
- Worldwide Distribution-- USA (nationwide) and the countries of Austria, Australia, Belgium, Brazil, Canada, China, Colombia, Germany, France, India, Israel, Italy, Japan, Korea, Mexico, The Netherlands, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, United Arab Emirate, and United Kingdom.
- Product codes and lots
- VTK23119 to VTK 23745