Guidant RAPIDO Cut-Away Guiding Catheter, 6F. Part Numbers: 66676-201 and 667552-100. For use in cardiac...
Abbott Vascular began a recall of Guidant RAPIDO Cut-Away Guiding Catheter, 6F. Part Numbers: 66676-201 and 667552-100. For use in cardiac diagnostic and therapeutic procedures on . The reason FDA records is "Five lots of Guide Catheters and Accessory products were processed through Ethylene Oxide (ETO) sterilization with additional plastic material covering the pallet, which is not consistent with validated packaging...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2793-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Vascular
- Product
- Guidant RAPIDO Cut-Away Guiding Catheter, 6F. Part Numbers: 66676-201 and 667552-100. For use in cardiac diagnostic and therapeutic procedures.
- Product code
- DQY Catheter, Percutaneous
- Reason for recall
- Five lots of Guide Catheters and Accessory products were processed through Ethylene Oxide (ETO) sterilization with additional plastic material covering the pallet, which is not consistent with validated packaging configuration for ETO sterilization.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Abbott Vascular sent an Urgent Device Recall letter, dated January 7, 2011, noting the products being recalled and the reasons for recall. The firm also provided the following reconciliation instructions to its territory manager: -Help customer identify aflected product. - CaIl Abbott Vascular Customer Service (800) 221-9902 to receive RGA number. Record RGA number below. - Fax this completed form to (951) 914-3826. - Return a copy of this completed form with the returned product
- Quantity in commerce
- 389 units total
- Distribution
- Nationwide distribution, and Ireland.
- Product codes and lots
- Lot Number: 0120891 and 0121091
FIRST ARTICLE / WEEKLY