Z-2677-2011
Remel Spectra MRSA, Ref R01821, 10 plates per package, and Ref R01822, 100 plates per package. The firm name...
Remel Inc / initiated 2011-05-11 / Terminated / root cause: manufacturing process
Remel Inc began a recall of Remel Spectra MRSA, Ref R01821, 10 plates per package, and Ref R01822, 100 plates per package. The firm name on the label is Remel, Lenexa, KS. The product is... on . The reason FDA records is "The product failed to prevent growth of methicillin-sensitive Staphylococcus aureus (MSSA)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2677-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Remel Inc
- Product
- Remel Spectra MRSA, Ref R01821, 10 plates per package, and Ref R01822, 100 plates per package. The firm name on the label is Remel, Lenexa, KS. The product is a selective and differential chromogenic medium recommended for use in the qualitative detection of nasal colonization of methicillin-resistant Staphylococcus aureus (MRSA) to aid in the prevention and control of MRSA in healthcare settings. It is also intended for use in the qualitative detection of MRSA from positive blood cultures demonstrating gram-positive cocci on Gram stain.
- Product code
- JSO Culture Media, Antimicrobial Susceptibility Test, Excluding Mueller Hinton Agar
- Reason for recall
- The product failed to prevent growth of methicillin-sensitive Staphylococcus aureus (MSSA)
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Remel, Inc. sent an "IMPORTANT MEDICAL DEVICE PRODUCT RECALL NOTICE" letter, dated May 10, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to review their inventory and discard all remaining product lots. Additionally, customers are to complete the Product Inventory Checklist and indicate whether or not they desire replacement of the product or credit. A self-addressed , stamped envelope was enclosed in the letter to use to return the checklist. Questions can be answered by contacting the Technical Services Department at 800-255-6730 or 913-888-0939.
- Quantity in commerce
- 187/10-packs and 261/100-packs
- Distribution
- Nationwide Distribution--including Puerto Rico.
- Product codes and lots
- Cat. # R01821 (10 pack) - Lot 028251, exp. 2011.06.13; Cat. #R01822 (100 pack) - Lot 028252, exp. 2011.06.13; 028883, exp. 2011.06.14; 028884, exp. 2011.06.14; and 031922, exp. 2011.06.21