Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2706-2011

Brand Name STAAR Surgical Collamer Ultraviolet-Absorbing Posterior Chamber Single Piece Foldable Intraocular...

Staar Surgical AG / initiated 2011-05-20 / Terminated

Staar Surgical Co. began a recall of Brand Name STAAR Surgical Collamer Ultraviolet-Absorbing Posterior Chamber Single Piece Foldable Intraocular Lens Common Names Collamer single piece IOL... on . The reason FDA records is "The recall was initiated by Staar Surgical due to an error in the Directions for Use (DFU) for Staar Surgical's Collamer Ultraviolet Absorbing Posterior Chamber Single Piece Foldable Intraocular Lens (Collamer IP IOL)...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2706-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Staar Surgical AG
Product
Brand Name STAAR Surgical Collamer Ultraviolet-Absorbing Posterior Chamber Single Piece Foldable Intraocular Lens Common Names Collamer single piece IOL Collamer plate haptic IOL The Device is intended to correct aphakia in persons 60 years of age or older in whom a cataractous lens has been removed by cataract extraction. The Device is to be implanted in the posterior chamber and in the capsular bag through a tear-free capsulorhexis (circular tear anterior capsulotomy).
Product code
HQL Intraocular Lens
Reason for recall
The recall was initiated by Staar Surgical due to an error in the Directions for Use (DFU) for Staar Surgical's Collamer Ultraviolet Absorbing Posterior Chamber Single Piece Foldable Intraocular Lens (Collamer IP IOL), also known as the nanoFLEX IOL. The Device Description erroneously indicates that the lens could be implanted in the ciliary sulcus.
FDA root cause
Other
Action
Staar Surgical send out an URGENT - Medical Device Correction Notice to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to acknowledge receipt of the letter by signing, dating, and faxing the form to Staar Surgical. For any questions call (800) 292-7902, ext 2345.
Quantity in commerce
16,750 devices distributed to the field
Distribution
Worldwide Distribution - USA (nationwide) and the countries of South Africa, Slovakia, Mexico, Hong Kong, and Sri Lanka
Product codes and lots
CC4204BF and CC4204A