Z-2709-2011

Product Brand Name: Versafitcup Double Mobility Liner Inserter Product Generic Name: Versafitcup Double...

Medacta USA, Inc. / initiated 2011-05-23 / Terminated / root cause: labeling and UDI

Medacta Usa Inc began a recall of Product Brand Name: Versafitcup Double Mobility Liner Inserter Product Generic Name: Versafitcup Double Mobility Liner Inserter Model Number: 01.26.10.0018 The... on . The reason FDA records is "Medacta International has recently realized that the base of the Liner Inserter Ref 01.26.10.0018 Lot 105549 of the Versafitcup Double Mobility system could become rusty after the washing procedures. This is due to an...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2709-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medacta USA, Inc.
Product
Product Brand Name: Versafitcup Double Mobility Liner Inserter Product Generic Name: Versafitcup Double Mobility Liner Inserter Model Number: 01.26.10.0018 The hip prosthesis is designed for cementless use in total hip arthroplasty in primary or revision surgery. The patient should be skeletally mature. The patients condition should be due to one or more of: " Severely painful and/or disabled joint: as a result of osteoarthritis, post-traumatic arthritis, rheumatoid arthritis or psoriactic arthritis, Congenital hip dysplasia, Ankylosing spondylitis. " Avascular necrosis of the femoral head. " Acute traumatic fracture of the femoral head or neck. " Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present. " Dislocation risks.
Product code
MEH Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate
Reason for recall
Medacta International has recently realized that the base of the Liner Inserter Ref 01.26.10.0018 Lot 105549 of the Versafitcup Double Mobility system could become rusty after the washing procedures. This is due to an incorrect raw material used by one of our suppliers during the production phase of this lot. For this reason, Medacta International has made the decision to remove all the pieces
Root cause tag
labeling and UDI
FDA root cause
Labeling False and Misleading
Action
Medacta USA sent an "URGENT SAFETY NOTIFICATION-VERSAFITCUP DM LINER INSERTER" letter dated May 23, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to check their inventory for the affected device, immediately sequester it and not use the device in surgery. The letter instructs customers to pack the affected device for shipment to Medacta USA under Return Material Authorization 1590. Customer will receive a replacement within a few days. Questions regarding this recall are directed to the Director of Regulatory and Quality at 805-437-7085 ext. 26.
Quantity in commerce
Worldwide: 18 pieces. In the USA: 9 pieces
Distribution
Worldwide Distribution--USA (nationwide) including the states of CA, ID, IL, and OH, and 9 products to foreign countries.
Product codes and lots
Lot code: 105549