Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2612-2011

45cm Laparoscopic Instruments Sterilization tray. Manufactured by: Stryker Endoscopy, 5900 Optical Court,San...

Stryker Corp. / initiated 2011-06-08 / Terminated / root cause: manufacturing process

Stryker Endoscopy began a recall of 45cm Laparoscopic Instruments Sterilization tray. Manufactured by: Stryker Endoscopy, 5900 Optical Court,San Jose, CA 95138 Designed to hold laparoscopic... on . The reason FDA records is "During testing and review, Stryker Endoscopy discovered that the sterilization parameters for the 132 degree C gravity steam sterilization cycle in the IFU may not consistently sterilize all instruments in the 45cm...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2612-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
45cm Laparoscopic Instruments Sterilization tray. Manufactured by: Stryker Endoscopy, 5900 Optical Court,San Jose, CA 95138 Designed to hold laparoscopic surgical manual instruments which are intended for cutting, holding, grasping and manipulating tissue, organs, organ area, surgical auxiliaries such as suturing material. The 45cm Lap tray can hold up to 11 instruments in the top tier and a combination of instruments in the bottom tier provided that the total weight of the loaded tray does not exceed 6.53 kg.
Product code
KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories
Reason for recall
During testing and review, Stryker Endoscopy discovered that the sterilization parameters for the 132 degree C gravity steam sterilization cycle in the IFU may not consistently sterilize all instruments in the 45cm Laparoscopic Instruments Tray.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Stryker Endoscopy sent a "URGENT: DEVICE CORRECTION" letter dated June 8, 2011 to affected customers. The letter describes the product, problem, and actions to be taken by the customers. Customers were instructed to discard all old Instructions for Use sheets for their 45cm Laparoscopic Tray and download the new IFU from the URL link:http://stryker.com/45cm tray. Additionally, customers were instructed to complete and return an attached "acknowledgement of receipt" form and fax to 408-754-8378 or scan and e-mail to [email protected]. Questions can be answered by calling 1-800-624-4422 or by e-mail at [email protected].
Quantity in commerce
603
Distribution
Worldwide Distribution--USA (nationwide) and countries of Brazil, Canada, Chili, Germany , Korea, Mexico, Philippines, Poland, Portugal, Singapore, Sweden, Switzerland, and Taiwan.
Product codes and lots
Model number 250-015-704. All serial/lot numbers.