Philips brand BrightView Gamma Camera System, SYST, BRIGHTVIEW-TILT, 3/4", Model Number: 2170-3003A; Product...

Philips North America LLC / initiated 2011-06-03 / Terminated / root cause: design

Philips Medical Systems began a recall of Philips brand BrightView Gamma Camera System, SYST, BRIGHTVIEW-TILT, 3/4", Model Number: 2170-3003A; Product is manufactured and distributed by Philips Medical... on . The reason FDA records is "The gap between the two parts of the table may permit pinches to occur on either patient's skin or other body parts; it has the potential for serious injury if a patient's finger is caught between the pallet and patient...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2869-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips brand BrightView Gamma Camera System, SYST, BRIGHTVIEW-TILT, 3/4", Model Number: 2170-3003A; Product is manufactured and distributed by Philips Medical Systems (Cleveland), Inc., Cleveland, OH The BrightView Gamma Camera System is intended to produce images depicting the anatomical distributions of single photon emitting radioisotopes within the human body for interpretation by medical personnel.
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
The gap between the two parts of the table may permit pinches to occur on either patient's skin or other body parts; it has the potential for serious injury if a patient's finger is caught between the pallet and patient table resulting in broken finger bone.
Root cause tag
design
FDA root cause
Device Design
Action
Philips Healthcare issued a Customer Information letter (Field Change Order (#88200402)) dated June 15, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to follow the instruction for use for their system, which states "Monitor the patient and the system vigilantly during any system motions to make sure that ...accessories, equipment, and all ojbects such as hair, jewelry, or neckties-remain clear of any moving camera and imaging table parts. For additional information, customers were istructed to see the BrightView SPECT instructions for Use, section 2.2 - Safety Warnings and Precautions and Safety and Information Labels. Philips Healthcare will notify all affected customers and will install a sticker through an Action for Performance - Proactive Field Change Order (FCO) to resolve the issue. For customers in North America who require further information or support contact Customer Care Solutions Center at 1-800-722-9377, Select Option 5. In all other countries the local Philips Healthcare office should be contacted.
Distribution
Worldwide Distribution - USA (nationwide) and Puerto Rico and the countries of Australia, Austria, Bahrain, Belgium, Brazil, China, Chile, Columbia, Czech Republic, Denmark, Dominican Republic, Egypt, France, Greece, India, Iran, Israel, Italy, Japan, Lithuania, Malaysia, Mexico, Netherlands, Peru, Portugal, Russia, Saudi Arabia, Singapore, South Korea, Spain, Switxerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, and the United Kingdom.
Product codes and lots
System Code: 882480;  Serial #s: 4000224, 4000061, 4000321, 4000043, 4000065, 4000081, 4000180, 4000044, 4000527, 4000298, 4000229, 4000262, 4000378, 4000280, 4000336, 4000324, 4000184, 4000190, 4000516, 4000303, 4000411, 4000305, 4000009, 4000012, 4000413, 4000213, 4000335, 4000436, 4000500, 4000491, 4000510, 4000518, 4000418, 4000454, 4000197, 4000353, 4000469, 4000005, 4000042, 4000379, 4000533, 4000451, 4000137, 4000343, 4000030, 4000540, 4000421, 4000342, 4000495, 4000511, 4000317, 4000338, 4000316, 4000408, 4000251, 4000031, 4000341, 4000122, 4000530, 4000395, 4000286, 4000294, 4000348, 4000296, 4000462, 4000463, 4000464, 4000419, 4000096, 4000476, 4340005, 4000484, 4000268, 4000382, 4000169, 4000447, 4000228, 4000442, 4000297, 4000047, 4000049, 4000488, 4000156, 4000412, 4000273, 4000337, 4000394, 4000194, 4000179, 4000401, 4000385, 4000504, 4000121, 4000109, 4000115, 4000257, 4000503, 4000105, 4000356, 4000333, 4000339, 4000107, 4000227, 4000233, 4000173, 4000400, 4000397, 4000398, 4000494, 4000256, 4000386, 4000168, 4000126, 4000154, 4000399, 4000272, 4000478, 4340004, 4000219, 4000243, 4000437, 4000376, 4000416, 4000485, 4000035, 4000515, 4000427, 4000142, 4000226, 4000519, 4000475 4000066, 4000183, 4000459, 4000461, 4000468, 4000210, 4000063 4000150, 4000217, 4000393, 4000420, 4000479, 4000300, 4000075, 4000080, 4000086, 4000013, 4000396, 4000481, 4000392, 4000157, 4000323, 4000374, 4000072, 4000237, 4000493, 4000354, 4000203, 4000167, 4000041, 4000407, 4000372, 4000133, 4000209, 4000299, 4000477, 4000314, 4000010, 4000539, 4000409, 4000134, 4000153, 4000373, 4000471, 4000163, 4000465, 4000082, 4000331, 4000282, 4000449, 4000216, 4000269, 4000415, 4000345, 4000208, 4000279, 4000254, 4000125, 4000201,