Calloway LabsMulti-Drug 9 Panel Assay A rapid, one step screening test for the simultaneous for the...
Alere San Diego began a recall of Calloway LabsMulti-Drug 9 Panel Assay A rapid, one step screening test for the simultaneous for the qualitative detection of multiple drugs and drug... on . The reason FDA records is "Alere is initiating a voluntary recall on the listed lots of MULTI-DRUG 9 PANEL INTEGRATED CUP manufactured by Alere for sole use by Calloway Labs due to a mistake that occurred while processing the order, these...". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.
Recall record
- Recall number
- Z-2881-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alere San Diego, Inc.
- Product
- Calloway LabsMulti-Drug 9 Panel Assay A rapid, one step screening test for the simultaneous for the qualitative detection of multiple drugs and drug metabolites in human urine.
- Product code
- DKZ Enzyme Immunoassay, Amphetamine
- Reason for recall
- Alere is initiating a voluntary recall on the listed lots of MULTI-DRUG 9 PANEL INTEGRATED CUP manufactured by Alere for sole use by Calloway Labs due to a mistake that occurred while processing the order, these products were incorrectly labeled as CLIA Waived, when in fact the devices are classified as Moderately Complex. Therefore these devices are misbranded.
- Root cause tag
- supplier
- FDA root cause
- Vendor change control
- Action
- Alere San Diego sent an "Urgent Medical Device Recall" letter dated June 9, 2011 to all affected customers. The letter included: products affected, reason for recall and instructed them to return all affected product. An acknowledgement form was also included for the customer to return to Alere. For questions contact Technical Services at 888.246.7483, option 2 or by e-mail at [email protected].
- Quantity in commerce
- 2400 kits (60,000 devices)
- Distribution
- (USA) Nationwide Distribution including the states of MO and MA.
- Product codes and lots
- Model/Catalog Number: DOA-197-201-19-0NW506A, Lot/Unit Codes: DOA1040725, DOA1040758, DOA1040784, DOA1040669 & DOA1040689