XiO Radiation Treatment Planning System, XiO versions equal to or greater than 2.4.0 and less than 4.63.0 The...

Elekta Limited / initiated 2011-06-09 / Terminated / root cause: software

Computerized Medical Systems Inc began a recall of XiO Radiation Treatment Planning System, XiO versions equal to or greater than 2.4.0 and less than 4.63.0 The XiO RTP System is used to create treatment plans... on . The reason FDA records is "The XiO Software currently prompts for entry of CT to ED data for a studyset. No indication is given that the Electron Density data should only be used for Photon and Electron calculations and either Relative Stopping...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2919-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
XiO Radiation Treatment Planning System, XiO versions equal to or greater than 2.4.0 and less than 4.63.0 The XiO RTP System is used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed.
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
The XiO Software currently prompts for entry of CT to ED data for a studyset. No indication is given that the Electron Density data should only be used for Photon and Electron calculations and either Relative Stopping Power or Mass Density should be used for Proton calculations (depending on the algorithm).
Root cause tag
software
FDA root cause
Software design
Action
The firm, Eletka, Inc., issued an "Important Safety Notice" dated May 18, 2011 via mail to all affected sites. The notice explains the issue and identifies a work around to the problem until the new release is available. The customers were instructed to complete and return the enclosed "Confirmation of Receipt" form via FAX to 314-993-1175, attn: QA; e-mail or mail: CMS Software, Attn: QA, 13723 Riverport Drive, Suite 100, Maryland Heights, MO 63043. The notice states that "this problem has existed since XiO Release 2.4.0 and has been resolved in XiO Release 4.63.0 by adding warning messages and explanations to the XiO On-Line Help. Your site will be notified when this release is available." If you have any question, call 314-993-0003, 800-878-4267 or 408-830-8023.
Quantity in commerce
7
Distribution
Worldwide distribution: USA (nationwide) including: IL, IN, MA, NJ and OK; and country of Germany.
Product codes and lots
XiO versions equal to or greater than 2.4.0 and less than 4.63.0