ARTISTE/RT Therapist v4.2 and/or Oncologist v4.2 systems. Manufactured by Siemens AG, Medical Solutions...
Siemens Medical Solutions USA, Inc began a recall of ARTISTE/RT Therapist v4.2 and/or Oncologist v4.2 systems. Manufactured by Siemens AG, Medical Solutions, Kemnath, Germany. Delivery of X-ray photon and... on . The reason FDA records is "A potential safety rick related to transferring reference images for offset calculation between a syngo RT Therapist v 4.2 or a syngo RT Oncologist 4.2 and other versions of the COHERENCE/syn go RT Therapist or...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2977-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- ARTISTE/RT Therapist v4.2 and/or Oncologist v4.2 systems. Manufactured by Siemens AG, Medical Solutions, Kemnath, Germany. Delivery of X-ray photon and electron radiation for the therapeutic treatment of cancer.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- A potential safety rick related to transferring reference images for offset calculation between a syngo RT Therapist v 4.2 or a syngo RT Oncologist 4.2 and other versions of the COHERENCE/syn go RT Therapist or Oncologist or vice versa. An incorrect offset may be applied for patient positioning and result in dose being delivered to the wrong location.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens sent an "URGENT: MEDICAL DEVICE CORRECTION/CUSTOMER SAFETY ADVISORY NOTICE" letter dated June 3, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter provides preventive measure instructions for customers to use until an update for the system is available.
- Quantity in commerce
- 121 devices
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of CA, FL, NC, NY, OH, PA, SD, UT, and WI and countries of Australia, Belgium, Croatia, Czech Republic, Germany, India, Ireland, Japan, Poland, Saudi Arabia, South Africa, Spain, Sweden, and United Kingdom.
- Product codes and lots
- The ARTISTE RT Therapist v4.2 and Oncologist v4.2 is used on the following: syngo RT Oncologist Part # 10652131; Therapist Express Basic, Part # 08151289; syngo RT Therapist, Part # 08162815; syngo RT Therapist Connect / MOSAIQ OSI, part # 08168754.