Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-3010-2011

Stryker Evacuation Chair Model 6254 It Is intended to aid in the transportation of passengers up and/or down...

Stryker Medical / initiated 2011-06-28 / Terminated / root cause: manufacturing process

Stryker Medical Division of Stryker Corporation began a recall of Stryker Evacuation Chair Model 6254 It Is intended to aid in the transportation of passengers up and/or down a flight of stairs in an evacuation situation on . The reason FDA records is "Some of the red release bars used during the time frame of 22 December 2010 through 31 January 2011 were reworked. The rework subjected the parts to excessive heat and as a result, the bars can bend during the action of...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-3010-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Medical
Product
Stryker Evacuation Chair Model 6254 It Is intended to aid in the transportation of passengers up and/or down a flight of stairs in an evacuation situation.
Product code
FPP Stretcher, Hand-Carried
Reason for recall
Some of the red release bars used during the time frame of 22 December 2010 through 31 January 2011 were reworked. The rework subjected the parts to excessive heat and as a result, the bars can bend during the action of squeezing the release bar against the black cross tube to release the track. The red release bars may bend to the degree that they permanently engage the track locking mechanism
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm, Stryker Medical, sent an "Urgent-Medical Device Recall" letter dated June 28, 2011 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately locate the listed product; inspect each unit prior to use; return the enclosed post card to confirm receipt of this notification; if they have loaned or sold any of the chairs listed, forward a copy this letter to the new users and advise Stryker of new users names and locations, and if they have disposed of any of the affected chairs and they are no longer in use, advise Stryker of their obsolescence by providing them with the serial numbers. Stryker stated in the letter that, their goal is the timely upgrade of all affected units within the next 90 days. If you have any questions or concerns, please contact us at 800-STRYKER, option 2. Please reference RA-2011-027. Normal business hours are Monday-Friday 8am-5pm EST.
Quantity in commerce
220
Distribution
Worldwide distribution: USA (nationwide) and countries of: Canada, China, France, Germany, India, Netherlands, and Sweden.
Product codes and lots
Model 6254, Lot numbers 110143822, 110143823, 110143824, 110143825, 110143826, 110143827, 101240120, 101240121, 101240122, 110142843, 110143758, 101240123, 101240124, 101240125, 101240126, 101240441, 101240442, 101240443, 101240863, 110142646, 110142632, 110142619, 110143756, 101240880, 110143801, 101240850, 101240851, 101240852, 101240853, 101240854, 101240855, 110143828, 110143829, 110143830, 110143831, 101241171, 101240127, 101240128, 101240129, 101240130, 101240131, 101240868, 101240892, 101240893, 101240894, 101240895, 101240896, 101240897, 101240898, 101240119, 110142620, 110143282, 110143283, 110143284, 110143281, 101240132, 101240133, 101240134, 101240135, 101240136, 101240137, 101240138, 101240139, 101240140, 101240141, 101240142, 101240143, 101240144, 101240858, 101240859, 101240860, 101240861, 101240862, 110143753, 110143754, 110143755, 101239046, 101239047, 101239048, 110142616, 110142617, 110142618, 101240145, 101240146, 110142625, 101241172, 101241173, 101241174, 101241175, 101241176, 101241177, 110142624, 101240870, 110142652, 110142653, 110142648, 110142649, 110142650, 110142651, 101241444, 101241445, 101241446, 101241447, 101241448, 101241449, 101240437, 101240438, 101240439, 101240440, 101241863, 101241864, 101241865, 101241866, 101240864, 101240865, 101240866, 101240867, 110142647, 110143222, 110142633, 110142634, 110142844, 110142621, 101240100, 101240101, 101240102, 101240103, 101240104, 101240105, 101240106, 101240107, 101240108, 101240109, 101240110, 101240111, 101240112, 101240113, 101240114, 101240881, 101240882, 101240883, 101240884, 110143832, 110143835, 110143797, 110143798, 110143799, 110143800, 110143750, 110143751, 110143757, 101240869, 110143802, 110143803, 110143804, 110143793, 110143794, 110143795, 110142630, 110142631, 101239030, 101239031, 101239032, 101239033, 101239034, 101239035, 101239036, 101239037, 101239038, 101239039, 101239040, 101239041, 101239042, 101239043, 101239044, 101239045, 101240416, 101240417, 101240418, 101240419, 101240420, 101240421, 101240422, 101240423, 101240424, 101240425, 101240426, 101240427, 101240428, 101240429, 101240430 01240431, 101240432, 101240433, 101240434, 101240435, 101241170, 101241152, 101240838, 101240839, 101240840, 101240841, 101240842, 101240843, 101240844, 101240845, 101240846, 101240408, 110143796, 101240118, 110142840, 110142841, 110142842, 101240409, 101240410, 101240411, 101240412, 101240413, 101240414 and 101240415.