Site Rite Vision Ultrasound System, REF 9770032, Bard Access Systems, Inc. Intended for diagnostic ultrasound...
Bard Access Systems began a recall of Site Rite Vision Ultrasound System, REF 9770032, Bard Access Systems, Inc. Intended for diagnostic ultrasound imaging or fluid flow analysis of the human body on . The reason FDA records is "Purchasing erroneously ordered the incorrect fan using the distributor's part number instead of using the specified manufacturer part number". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2782-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bard Access Systems, Inc.
- Product
- Site Rite Vision Ultrasound System, REF 9770032, Bard Access Systems, Inc. Intended for diagnostic ultrasound imaging or fluid flow analysis of the human body.
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- Purchasing erroneously ordered the incorrect fan using the distributor's part number instead of using the specified manufacturer part number.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Bard Access Systems, Inc sent out an "Urgent" Medical Device Recall (User Level)" dated June 2011 to all affected customers The letter included: product name, description of problem and requested the return of units bearing "B50" on the fan. Bard Access Systems, Inc. will replace the incorrect fan with the correct fan. For additional information call 1-724-779-3000.
- Quantity in commerce
- 86 units
- Distribution
- Nationwide (USA) distribution including the states of AL, AR, AZ, CA, CO, FL, IA, IN, MD, MI, MO, MS, NC, NJ, NM, NY, OH, PA, RI, SC, TX, UT, VA, VT, and WA.
- Product codes and lots
- Serial numbers: DYUJA023, DYUJA024, DYUJA025, DYUJA026, DYUJA027, DYUJA028, DYUJA029, DYUJA030, DYUJA031, DYUJA032, DYUJA034, DYUJA036, DYUJA037, DYUJA038, DYUJA039, DYUJA040, DYUKA001, DYUKA002, DYUKA003, DYUKA004, DYUKA005, DYUKA006, DYUKA007, DYUKA008, DYUKA009, DYUKA010, DYUKA011, DYUKA012, DYUKA013, DYUKA014, DYUKA015, DYUKA015, DYUKA017, DYUKA018, DYUKA019, DYUKA020, DYVAA001, DYVAA002, DYVAA003, DYVAA004, DYVAA005, DYVAA006, DYVAA007, DYVAA012, DYVAA013, DYVAA014, DYVAA015, DYVAA016, DYVAA017, DYVAA018, DYVAA019, DYVAA020, DYVAA021, DYVAA022, DYVAA023, DYVAA024, DYVAA025, DYVAA026, DYVAA027, DYVAA028, DYVAA029, DYVAA030, DYVAA031, DYVAA032, DYVAA033, DYVAA036, DYVAA037, DYVAA038, DYVAA039, DYVAA040, DYVBA002, DYVBA003, DYVBA004, DYVBA005, DYVBA006, DYVBA007, DYVBA008, DYVBA009, DYVBA010, DYVBA011, DYVBA012, DYVBA013, DYVBA014, DYVBA015, DYVBA018, DYVBA019, DYVBA020.