Z-2894-2011

Arrow ASK-20608-CMC StimuCath Continuous Nerve Block Kit. Arrow International, Inc. Permits placement of...

Arrow International LLC (subsidiary of Teleflex Incorporated) / initiated 2011-06-17 / Terminated / root cause: packaging

Arrow International Inc began a recall of Arrow ASK-20608-CMC StimuCath Continuous Nerve Block Kit. Arrow International, Inc. Permits placement of catheters next to nerves and nerve plexuses for... on . The reason FDA records is "Some of the outer trays used in the packaging were found to have an insufficient seal. Therefore, packaging may have been compromised, and sterility of the product cannot be guaranteed". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2894-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Arrow International LLC (subsidiary of Teleflex Incorporated)
Product
Arrow ASK-20608-CMC StimuCath Continuous Nerve Block Kit. Arrow International, Inc. Permits placement of catheters next to nerves and nerve plexuses for continuous nerve block anesthesia or analgesia techniques for periods not exceeding 72 hours.
Product code
CAZ Anesthesia Conduction Kit
Reason for recall
Some of the outer trays used in the packaging were found to have an insufficient seal. Therefore, packaging may have been compromised, and sterility of the product cannot be guaranteed.
Root cause tag
packaging
FDA root cause
Packaging
Action
Arrow International, Inc. sent an "Urgent Medical Device Recall" letter dated June 17, 2011 to all affected customers. The letter describes the problem and provides recommendations including; quarantine affected product, return of product, and send an acknowledgement form back to manufacturer. For additional information call 800 233-3187.
Quantity in commerce
95 kits
Distribution
Nationwide Distribution including the state of North Carolina.
Product codes and lots
Lot # MF0126544