Z-0128-2012

SYNCHRON CX5 Delta Clinical System Part #: 468243

Siemens Healthcare Diagnostics Inc. / initiated 2010-08-23 / Terminated / root cause: software

Beckman Coulter Inc. began a recall of SYNCHRON CX5 Delta Clinical System Part #: 468243 Product Usage: The Beckman SYNCHRON CX4 DELTA, CX7 DELTA, CX4CE, and CX7 systems are fully automated and... on . The reason FDA records is "The recall was initiated because Beckman Coulter has confirmed software limitations which affect only manually-loaded reagent cartridges on the CX4/5/7 CE, CX4/5/7 Delta and CX9 ALX Clinical Systems". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0128-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
SYNCHRON CX5 Delta Clinical System Part #: 468243 Product Usage: The Beckman SYNCHRON CX4 DELTA, CX7 DELTA, CX4CE, and CX7 systems are fully automated and computer controlled instruments designed for the in vitro diagnostic quantitation of biological fluid components and therapeutic drugs as well as the qualitative determination of drugs of abuse in urine
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
The recall was initiated because Beckman Coulter has confirmed software limitations which affect only manually-loaded reagent cartridges on the CX4/5/7 CE, CX4/5/7 Delta and CX9 ALX Clinical Systems.
Root cause tag
software
FDA root cause
Software design
Action
Beckman Coulter sent an "URGENT: PRODUCT CORRECTIVE ACTION" letter dated September 20, 2010 with an attached Fax Back Response Form (via US Postal Service for US Customers) to all customers who purchased the Synchron CX4/5/7 CE, CX4/5/7 Delta and CX9 ALX Clinical Systems. The letter provides the customers wtih an explanation of the problem identified and actions to be taken. Customers were instructed to share this information with laboratory staff and retain this notification as part of their laboratory Quality System documentation. If they forwarded the affected product listed above to another laboratory, they were asked to provide a copy of this letter to them. Questions regarding this Product Corrective Action letter, were directed to Customer Support Center (Call Center) on the Beckman Coulter website at http://www.beckmancoulter.com/customersupport/support or call (800) 854-3633 in the United States and Canada. Outside the United States and Canada, they were directed to their local Beckman Coulter Representative.
Quantity in commerce
1070 systems total in US
Distribution
Worldwide Distribution - USA (nationwide) and the country of Canada
Product codes and lots
All non-PRO Software Versions