MAGNETOM Aera/Skyra System - Nuclear magnetic resonance imaging

Siemens Medical Solutions USA, Inc. / initiated 2011-06-20 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA, Inc began a recall of MAGNETOM Aera/Skyra System - Nuclear magnetic resonance imaging on . The reason FDA records is "Firm discovered during product monitoring that on a few systems the gradient cable connections did not meet their specifications after installation. Locally, the cable connections overheated, making servicing...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2979-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
MAGNETOM Aera/Skyra System - Nuclear magnetic resonance imaging
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
Firm discovered during product monitoring that on a few systems the gradient cable connections did not meet their specifications after installation. Locally, the cable connections overheated, making servicing necessary.The potential exists for a cable connection to become overheated, leading to the emission of smoke and possible material damage.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Siemens sent a "CUSTOMER SAFETY ADVISORY NOTICE" letter dated June 20, 2011 to all affected customers. The letter describes the product, problem, and actions that will be taken. The letter informs customers that an inspection of the gradient cable connections and an emergency shutdown mechanism will be installed by a field service representative during a visit to all affected locations.
Quantity in commerce
26
Distribution
Nationwide Distribution-USA (nationwide) including the states of AK, FL, GA, IL, MA, MI, MN, MT, NC, NJ, NY, OH, TX, UT, VA, WA, and WV.
Product codes and lots
Model numbers 10432914 and 10432915