Z-3192-2011

Sterile Bone Cutting Burrs model S-5B-4 labeled in part:***S-5B-4***QTY: 1***5 mm Fluted Ball...

The Anspach Effort, Inc. / initiated 2011-05-09 / Terminated / root cause: manufacturing process

The Anspach Effort, Inc. began a recall of Sterile Bone Cutting Burrs model S-5B-4 labeled in part:***S-5B-4***QTY: 1***5 mm Fluted Ball, 4-Series.***Use with***SHORT Attachments.***Use by... on . The reason FDA records is "The Anspach Effort, Inc., Palm Beach Gardens, FL, is recalling their Anspach Single Use, Sterile Bone Cutting Burrs, catalog numbers, IMRI-S-8NS-M, M-6P, QD8-4OVD, S-5B-4, and SIL-7-9, lot #'s C323023789, C403026551...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-3192-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
The Anspach Effort, Inc.
Product
Sterile Bone Cutting Burrs model S-5B-4 labeled in part:***S-5B-4***QTY: 1***5 mm Fluted Ball, 4-Series.***Use with***SHORT Attachments.***Use by 2016-03***ANSPACH***Manufacturer***Made in USA***The Anspach Effort, Inc.***4500 Riverside Drive***Palm Beach Gardens, FL 33410, USA***Tel: (800) 327 6887 / + 1 561 627 1080***EC REP***Anspach Europe Ltd.***Aston Court, Kingsmead Business Park***Frederick Place***High Wycombe, HP 11 1LA***United Kingdom***Tel: +441494 616126***Rx Only***LOT E173049623***. Cutting and Shaping bone including bones of the spine and cranium.
Product code
HBC Motor, Drill, Electric
Reason for recall
The Anspach Effort, Inc., Palm Beach Gardens, FL, is recalling their Anspach Single Use, Sterile Bone Cutting Burrs, catalog numbers, IMRI-S-8NS-M, M-6P, QD8-4OVD, S-5B-4, and SIL-7-9, lot #'s C323023789, C403026551, C473028567, C403026563, D153034013, D173034385, and C363024789, because the bar code which defines the expiration date may be incorrect. This could cause the bar code reader system to
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
The Anspach Effort, Inc. sent an "URGENT: MEDICAL DEVICE REMOVAL" letter dated May 27, 2011 to all affected customers. The letter described the product, problem, and the actions to be taken by the customers. The letter instructs customers to check their inventory and to segregate any affected product found. A Customer Response Form was attached for customers to complete and return to the firm via fax at 1-800-327-6661. Contact Anspach Customer Support at 800-327-6887 for questions regarding this recall.
Quantity in commerce
30
Distribution
Worldwide Distribution--USA (nationwide) including the states of Alabama, California, and Wisconsin and the countries of Columbia, France, Germany, Great Britain, Italy, Panama, Poland, Saudi Arabia, South Africa, Spain, and Switzerland.
Product codes and lots
Batch #s - D153034013, D173034385