Z-3028-2011
BD Vacutainer Push Button Blood Collection Set; REF 367324 Rx only, 23 G x 3/4" x 12 " (0.6 x 19 mm x 305...
Becton Dickinson & Co / initiated 2011-04-22 / Terminated / root cause: design
Becton Dickinson & Company began a recall of BD Vacutainer Push Button Blood Collection Set; REF 367324 Rx only, 23 G x 3/4" x 12 " (0.6 x 19 mm x 305 mm); BD Vacutainer Push Button Blood Collection Set... on . The reason FDA records is "BD Vacutainer Push Button Blood Collection Set can potentially undergo cannula separation from the hub". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-3028-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD Vacutainer Push Button Blood Collection Set; REF 367324 Rx only, 23 G x 3/4" x 12 " (0.6 x 19 mm x 305 mm); BD Vacutainer Push Button Blood Collection Set without Multiple Sample Luer Adaptor, Sterile, Rx only, 21 G x 3/4 x 12" 0.8 x 19 mm x 305 mm Blood collection system.
- Product code
- KST System, Blood Collection, Vacuum-Assisted, Manual
- Reason for recall
- BD Vacutainer Push Button Blood Collection Set can potentially undergo cannula separation from the hub.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- BD Diagnostics sent a "Lot Specific Product Recall" letter dated April 22, 2011 to all affected customers. The letter asked customers return any affected product and provides instructions for its return. The letter ask customers return the attached Packaging List with the shipment. For question on this recall contact BD Diagnostics at (201) 847-5117.
- Quantity in commerce
- 1,744,000
- Distribution
- Nationwide (USA) Distribution.
- Product codes and lots
- Catalog # 367324, Lot #0315173, 0316102, 0323342, 0327441, and 1025938; Catalog # 367326, Lot #0309244, 0309286, and 0312571; Catalog #367344, Lot #0334101, 0336557, 0347425, 0354635, and 10131090.