Z-3106-2011
Depuy Mitek Ratchet Handle with Quick Connect
DePuy Mitek, Inc. / initiated 2011-06-27 / Terminated
DePuy Mitek, Inc., a Johnson & Johnson Co. began a recall of Depuy Mitek Ratchet Handle with Quick Connect Product code: 219970. Instrument used with anchor inserters for insertion of DePuy Mitek Interference Screws... on . The reason FDA records is "Quick Disconnect Ratchet Handle may slip up the driver shaft or disengage leaving the shaft in the screw head". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-3106-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- DePuy Mitek, Inc.
- Product
- Depuy Mitek Ratchet Handle with Quick Connect Product code: 219970. Instrument used with anchor inserters for insertion of DePuy Mitek Interference Screws (orthopedic).
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Quick Disconnect Ratchet Handle may slip up the driver shaft or disengage leaving the shaft in the screw head.
- FDA root cause
- Process change control
- Action
- Depuy Mitek issued Original Customer Notification Letter on 6/27/11 and an Updated letter to clarify the lot number information sent 6/29/11. Hospitals requested to immediately check all inventories for affected product and return for repalcement product. Also notify surgeons aware of this potential issue and potential risk of a delay in surgery. questions or concerns with regard to this program, please contact your local DePuy Mitek Representative or DePuy Mitek Customer Service directly at 1-800-382-4682, PROMPT 3.
- Quantity in commerce
- 75 units
- Distribution
- Nationwide distribution.
- Product codes and lots
- Finished Goods (Box label) Lot Numbers: 1105101 and 1106132. Component lot number Etched on Reusable Handle 1012001 and 1106001