Access Immunoassay Systems Total T4 Calibrators, Part Number: 33805, Subsequent
Beckman Coulter Inc. began a recall of Access Immunoassay Systems Total T4 Calibrators, Part Number: 33805, Subsequent Product Code: JIS The Access Total T4 Calibrators are intended to calibrate the... on . The reason FDA records is "The recall was initiated because Beckman Coulter has confirmed that the stability claims for the Access Total T4 calibrators may not be met. This issue is associated with calibrator lots 917074 and 008534". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-3027-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Access Immunoassay Systems Total T4 Calibrators, Part Number: 33805, Subsequent Product Code: JIS The Access Total T4 Calibrators are intended to calibrate the Access Total T4 assay for the quantitative determination of total thyroxine (T4) levels in human serum and plasma using the Access Immunoassay Systems.
- Product code
- KLI Enzyme Immunoassay, Non-Radiolabeled, Total Thyroxine
- Reason for recall
- The recall was initiated because Beckman Coulter has confirmed that the stability claims for the Access Total T4 calibrators may not be met. This issue is associated with calibrator lots 917074 and 008534.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Beckman Coulter sent an "URGENT: PRODUCT CORRECTIVE ACTION" letter dated September 13, 2010 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to discontinue use of the affected lot numbers and review historical quality control performance. Additionally, a response form was attached to the letter for customers to complete and return. Alternate lots will be provided for current and future replenishment orders or requests. Customers with questions regarding this notification can call Technical Support at (800) 854-3633 or contact their local Beckman Coulter representative.
- Quantity in commerce
- 2495 units total (2492 in US)
- Distribution
- Worldwide Distribution-USA (nationwide) and the country of Canada.
- Product codes and lots
- Lot Number: 917074, 008534 Expiration Date: 917074 (12/31/10), 008534 (2/28/11)