PALL Medical LeukoGuard LGB Arterial Line Leukocyte Reduction Filter with Bypass Loop, STERILE...
Pall Corporation began a recall of PALL Medical LeukoGuard LGB Arterial Line Leukocyte Reduction Filter with Bypass Loop, STERILE. CATALOG/MODEL/Reorder Number: LGB. CLASSIFICATION NAME: Filter... on . The reason FDA records is "Potential for the housing of the arterial line Leukocyte Reduction Filter to separate before or during the priming procedure". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2856-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Pall Corporation
- Product
- PALL Medical LeukoGuard LGB Arterial Line Leukocyte Reduction Filter with Bypass Loop, STERILE. CATALOG/MODEL/Reorder Number: LGB. CLASSIFICATION NAME: Filter, Blood, Cardiopulmonary Bypass Arterial Line. COMMON/USUAL NAME: Arterial Line Filter with Bypass Loop. -- Manufactured by: PALL NEWQUAY, CORNWALL, TR9 6TT, UK -- Device Listing # D000745. --- Designed to reduce the levels of circulating leukocytes and remove microemboli greater than 40 microns in size from the perfusate during extracorporeal circulation including: gas emboli, fat emboli and aggregates composed of platelets, red blood cells and other debris.
- Product code
- LGB Gonococcal Antibody Tests
- Reason for recall
- Potential for the housing of the arterial line Leukocyte Reduction Filter to separate before or during the priming procedure.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Pall Medical sent an "Important Customer Notification" letter dated May 26, 2011 to all affected customers. The letter included a description of the problem, affected lots and provided recommendations including the return of the product. For additional information, please contact the firm at (516) 484-0263.
- Quantity in commerce
- 232 units
- Distribution
- Nationwide (USA) Distribution including the states of CA, LA, MI, TX and WI.
- Product codes and lots
- Catalog/Model/Reorder Number: LGB Lot # 10-405
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.